Skip to the content

FDA Medical Device Recalls (openFDA)

Shimadzu Medical Systems: R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1.2P324DK-125, 06/1.2P324DK-85, 0.6/1.2P38DE-85. Product Usage: The device is the beam-limiting device for X-ray Radiography.

Recall number
Z-0261-2013
Recalling firm
Shimadzu Medical Systems
Product
R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1.2P324DK-125, 06/1.2P324DK-85, 0.6/1.2P38DE-85. Product Usage: The device is the beam-limiting device for X-ray Radiography.
Status
Terminated
Initiated
2011-10-28
Posted
2012-11-08
Terminated
2016-07-26
Reason
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
Root cause
Nonconforming Material/Component
Action
Shimadzu sent an Urgent Voluntary Medical Device Recall - Field Safety Notice letters dated September 19, 2012 to all the affected customers. The letter identifed the description of non-compliance, how to identify the affected products, corrective action plan by Shimadzu and actions to be taken customers. Customers were instructed to contact their local Shimadzu representative or contact Shimadzu National Technical Support at Shimadzu Medical Systems National Technical Support (800) 228-1429.
Quantity
308 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
Device Model #R-20J
510(k) numbers
["K101036"]
PMA numbers
not on file
Product code
IZX
Firm FEI
1000125362
Firm city
Torrance
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register