FDA Medical Device Recalls (openFDA)
American Medical Systems, Inc.: AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI).
- Recall number
- Z-0259-2013
- Recalling firm
- American Medical Systems, Inc.
- Product
- AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI).
- Status
- Terminated
- Initiated
- 2012-10-19
- Posted
- 2012-11-07
- Terminated
- 2013-04-11
- Reason
- American Medical Systems (AMS) is issuing a voluntary recall on a number of lots of the AdVance Male Sling System (AdVance). AMS has recently received a small number of complaints regarding the resorbable suture in the AdVance sling not lasting for the entire 15 month shelf life of the product. An internal investigation has been conducted and confirmed that some sutures degraded prior to the
- Root cause
- Device Design
- Action
- American Medical Systems sent an "URGENT MEDICAL DEVICE RECALL" dated October 18, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter instructs customers to return any existing stock of the recall product and to complete and return the Product Recall Acknowledgment Form. Contact AMS Customer Service at 1-800-328-3881 for additional information or questions. .
- Quantity
- 4872
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of Austria, Australia, Brazil, Canada, Germany, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Croatia, India, Morocco, Mexico, Malaysia, Netherlands, New Zealand, Panama, Poland, Russian Federation, Solomon Islands, Singapore, Thailand, Turkey, Venezuela, South Africa, Chile, Colombia, Costa Rica, Slovenia, Argentina, China, and Taiwan.
- Lot, serial or model codes
- Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 790266012 72404224 742528001 to 795689010
- 510(k) numbers
- ["K053371"]
- PMA numbers
- not on file
- Product code
- FTL
- Firm FEI
- 1000116179
- Firm city
- Minnetonka
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28