Skip to the content

FDA Medical Device Recalls (openFDA)

DRG International, Inc.: 17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma Model/Catalog Number: EIA-1292

Recall number
Z-0258-2025
Recalling firm
DRG International, Inc.
Product
17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma Model/Catalog Number: EIA-1292
Status
Open, Classified
Initiated
2024-09-13
Posted
2024-10-29
Terminated
not on file
Reason
The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.
Root cause
Process change control
Action
DRG issued Urgent Field Safety Notification on 9/17/24 via email. Letter states reason for recall, health risk and action to take: Please check your inventory for the respective lot and ensure that the proper acceptance ranges provided below are used to validate the assay. Immediately reply to email: qa@drg-international.com to confirm this notice and your compliance.
Quantity
504 units
Distribution
AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam
Lot, serial or model codes
Lot Code: 04048474012927 (DRG Instruments GmbH) lots: 22K024 and 22K034
510(k) numbers
not on file
PMA numbers
not on file
Product code
JLX
Firm FEI
2245285
Firm city
Springfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register