AGA Medical Corporation: Amplatzer Cardiac Septal Occluder delivery system Exchange Systems labeled as follows: a) AMPLATZER Exchange System, Order no. 9-EXCH-6F-180/80, 9-EXCH-8F-180/80, 9EXCH-9F-45/80, 9-EXCH-12F-45/80. [Breakdown of order no. ''9-EXCH-6F180/80'' is as follows: 9-EXCH, French size (6F) - degree angle curve (180) / usable length in cm (80).] and b) the following devices are not approved for US distribution: AMPLATZER TorqVue Exchange System, Order no. 9-EITV09F45/80, 9-EITV12F45/80, 9-EITV06F180/80, 9-EITV08F180/80. Manufactured in USA, Sterile EO. Headquarters AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 U.S.A. The Exchange System is comprised of a delivery sheath, delivery cable, dilator, loading device and pin vise.
- Recall number
- Z-0256-2007
- Recalling firm
- AGA Medical Corporation
- Product
- Amplatzer Cardiac Septal Occluder delivery system Exchange Systems labeled as follows: a) AMPLATZER Exchange System, Order no. 9-EXCH-6F-180/80, 9-EXCH-8F-180/80, 9EXCH-9F-45/80, 9-EXCH-12F-45/80. [Breakdown of order no. ''9-EXCH-6F180/80'' is as follows: 9-EXCH, French size (6F) - degree angle curve (180) / usable length in cm (80).] and b) the following devices are not approved for US distribution: AMPLATZER TorqVue Exchange System, Order no. 9-EITV09F45/80, 9-EITV12F45/80, 9-EITV06F180/80, 9-EITV08F180/80. Manufactured in USA, Sterile EO. Headquarters AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 U.S.A. The Exchange System is comprised of a delivery sheath, delivery cable, dilator, loading device and pin vise.
- Status
- Terminated
- Initiated
- 2006-10-02
- Posted
- 2006-12-19
- Terminated
- 2008-10-09
- Reason
- AGA Medical has found that microscopic tears can occur in the delivery system sterile packaging under accelerated stress testing with routine shipping configurations. These microscopic tears are a potential breach of the sterile barrier. AGA has no confirmed complaints or adverse events related to this failure mode.
- Root cause
- Other
- Action
- An Urgent Recall Notice, dated 09/29/06, was sent to customers via facsimile or email and via Federal Express. The letter describes the issue and what product is affected. Unexpired product that is returned to AGA will be replaced. A response form is asked to be returned to AGA.
- Quantity
- 1,967
- Distribution
- Worldwide, including USA, Puerto Rico, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cuba, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Israel, Italy, Japan, Jordan, Korea, Kuwait, Malaysia, Moroc, Mexico, Norway, Pakistan, Peru, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, Slovak Republic, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela, and Vietnam.
- Lot, serial or model codes
- Order no. 9-EXCH6F-180/80, Lot Numbers: M05F10-13, M06F22-25. Order no. 9-EXCH8F-180/80, Lot numbers: M04D01-06, M04E27-21, M05C23-13. Order no. 9EXCH9F-45/80, Lot Numbers: M03L18-54, M05A12-09, M05D19-10, M05D18-47, M05E23-37, M05G15-06, M05H18-36. Order no. 9-EXCH12F-45/80, Lot Numbers: M05A13-23, M05B04-11, M05B14-08, M05C30-11, M05E02-19, M05H08-20. the following devices are not approved for US distribution: AMPLATZER TorqVue Exchange System, Order no. 9-EITV09F45/80, Lot numbers: M06A05-31, M05L15-24. Order no. 9-EITV12F45/80, Lot numbers: M06A05-33, M05L15-26. Order no. 9-EITV06F180/80, Lot numbers: M06D19-51. Order no. 9-EITV08F180/80, Lot numbers: M05L15-32, M06G26-02.
- 510(k) numbers
- not on file
- PMA numbers
- ["P020024"]
- Product code
- GBK
- Firm FEI
- 3005334138
- Firm city
- Golden Valley
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28