FDA Medical Device Recalls (openFDA)
Philips Medical Systems: SKYLight Gamma Camera System; SKYLIGHT,AZ,9FT,5/8 Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight Gamma Camera System is intended to product images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
- Recall number
- Z-0254-2012
- Recalling firm
- Philips Medical Systems
- Product
- SKYLight Gamma Camera System; SKYLIGHT,AZ,9FT,5/8 Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight Gamma Camera System is intended to product images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
- Status
- Terminated
- Initiated
- 2011-09-27
- Posted
- 2011-11-19
- Terminated
- 2012-12-21
- Reason
- A mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
- Root cause
- Device Design
- Action
- Philips Medical Systems sent a "Customer Information" letter dated September 27, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter lists additional precautions the customers should use while operating the device. Philips will notify customers for replacing the affected product through an Action for Performance-Proactive Field Change Order to resolve the issue. Contact Philips Customer Care Solutions at 1-800-722-9377, option 5, for questions concerning this notice.
- Quantity
- 658 - all units in recall
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico.
- Lot, serial or model codes
- Model number: 2160-3001B
- 510(k) numbers
- ["K000908"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 1525965
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28