Biocare Medical, LLC: Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.0 ml prediluted and ready to use. They are the same product with the same intended use. Chemistry - IVD Smooth Muscle Actin (SMA) antibody recognizes the alpha-smooth muscle isoform of actin. This antibody recognizes the alpha-smooth muscle isoform of actin. This MAb is reportedly useful for identifying tumors arising from smooth muscle and myoepithelial cells.
- Recall number
- Z-0252-2016
- Recalling firm
- Biocare Medical, LLC
- Product
- Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.0 ml prediluted and ready to use. They are the same product with the same intended use. Chemistry - IVD Smooth Muscle Actin (SMA) antibody recognizes the alpha-smooth muscle isoform of actin. This antibody recognizes the alpha-smooth muscle isoform of actin. This MAb is reportedly useful for identifying tumors arising from smooth muscle and myoepithelial cells.
- Status
- Terminated
- Initiated
- 2015-09-08
- Posted
- 2015-11-09
- Terminated
- 2016-01-11
- Reason
- Certain catalog numbers and lots of Smooth Muscle Actin may stain some cell types that are typically expected to be negative for SMA.
- Root cause
- Device Design
- Action
- Biocare sent a Medical Device Recall Notification letter dated September 8, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue and dispose of all affected product. Customers were asked to complete an return the enclosed Return Response Form. Customer's accounts would be credited for all affected product. For questions regarding this recall call 925-603-8033.
- Quantity
- 151 units
- Distribution
- Worldwide Distribution - US including CO, FL, OH, SC, AZ, AL, FL, CT, NJ, CA, DC, KY, NC, TX, ID, MA, MD, OR, LA, PA, KS, NY, UT, GA and Internationally to Spain, Sweden, Switzerland, Hungary, Italy, Germany, Turkey, Canada, India, Bolivia, Mexico, Indonesia and Australia.
- Lot, serial or model codes
- Catalog numbers: CME305A & CME305B: lots 110112, Expiry 11/2015;0 42213, expiry 4/2016; 101013, expiry 10/2016; 121813, expiry 12/2016; 070214, expiry 7/2017; 020215, expiry 2/2018. Catalog number: PME305AA, lots 070513, expiry 07/2016; 032114, expiry 3/2017, 030315, expiry 3/2018.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NJT
- Firm FEI
- 3004140393
- Firm city
- Concord
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28