Skip to the content

FDA Medical Device Recalls (openFDA)

Biocare Medical, LLC: Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.0 ml prediluted and ready to use. They are the same product with the same intended use. Chemistry - IVD Smooth Muscle Actin (SMA) antibody recognizes the alpha-smooth muscle isoform of actin. This antibody recognizes the alpha-smooth muscle isoform of actin. This MAb is reportedly useful for identifying tumors arising from smooth muscle and myoepithelial cells.

Recall number
Z-0252-2016
Recalling firm
Biocare Medical, LLC
Product
Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.0 ml prediluted and ready to use. They are the same product with the same intended use. Chemistry - IVD Smooth Muscle Actin (SMA) antibody recognizes the alpha-smooth muscle isoform of actin. This antibody recognizes the alpha-smooth muscle isoform of actin. This MAb is reportedly useful for identifying tumors arising from smooth muscle and myoepithelial cells.
Status
Terminated
Initiated
2015-09-08
Posted
2015-11-09
Terminated
2016-01-11
Reason
Certain catalog numbers and lots of Smooth Muscle Actin may stain some cell types that are typically expected to be negative for SMA.
Root cause
Device Design
Action
Biocare sent a Medical Device Recall Notification letter dated September 8, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue and dispose of all affected product. Customers were asked to complete an return the enclosed Return Response Form. Customer's accounts would be credited for all affected product. For questions regarding this recall call 925-603-8033.
Quantity
151 units
Distribution
Worldwide Distribution - US including CO, FL, OH, SC, AZ, AL, FL, CT, NJ, CA, DC, KY, NC, TX, ID, MA, MD, OR, LA, PA, KS, NY, UT, GA and Internationally to Spain, Sweden, Switzerland, Hungary, Italy, Germany, Turkey, Canada, India, Bolivia, Mexico, Indonesia and Australia.
Lot, serial or model codes
Catalog numbers:  CME305A & CME305B: lots 110112, Expiry 11/2015;0 42213, expiry 4/2016; 101013, expiry 10/2016; 121813, expiry 12/2016; 070214, expiry 7/2017; 020215, expiry 2/2018.  Catalog number: PME305AA, lots 070513, expiry 07/2016; 032114, expiry 3/2017, 030315, expiry 3/2018.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NJT
Firm FEI
3004140393
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register