FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics, MIS MODULAR DISTAL CAPTURE TRIATHLON MIS INSTRUMENTS, REF 6541-5-723, Lot AdminECR19205, Non-Sterile, The Triathlon MIS Distal Capture is used when a surgeon elects to use a capture for the distal femoral resection in a Triathlon primary MIS TKA. Once the Triathlon MIS distal Resection Guide (cat. no. 6541-5-721/722, for left or right preparation, respectively) has been secured to the femoral, the MIS distal capture can optionally be attached to the distal resection guide to complete the cut, per Trithion MIS surgical protocol TRI ATH-SP-5. Surgical instrument for knee prosthesis implantation.
- Recall number
- Z-0249-2016
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Stryker Orthopaedics, MIS MODULAR DISTAL CAPTURE TRIATHLON MIS INSTRUMENTS, REF 6541-5-723, Lot AdminECR19205, Non-Sterile, The Triathlon MIS Distal Capture is used when a surgeon elects to use a capture for the distal femoral resection in a Triathlon primary MIS TKA. Once the Triathlon MIS distal Resection Guide (cat. no. 6541-5-721/722, for left or right preparation, respectively) has been secured to the femoral, the MIS distal capture can optionally be attached to the distal resection guide to complete the cut, per Trithion MIS surgical protocol TRI ATH-SP-5. Surgical instrument for knee prosthesis implantation.
- Status
- Terminated
- Initiated
- 2015-09-29
- Posted
- 2015-11-09
- Terminated
- 2016-09-23
- Reason
- Stryker is recalling various lots of the Trithion MIS Modular Distal Capture because complaints were received that specific lots of this product may disassociate during use.
- Root cause
- Nonconforming Material/Component
- Action
- Stryker notified their Branches/Agencies of this recall by email on September 29, 2015 and notifiction letters and product accountability forms were sent via UPS (with return receipt) on September 29, 2015 to branches and on September 30, 2015 to hospitals. Stericycle wil be handling the returns. Customers with questions were instructed to call 201-831-5826. For questions regarding this recall call 201-831-5272.
- Quantity
- 3,444 units
- Distribution
- Nationwide Distribution including AZ, DE, ID, IL, KS, LA, MI, NH, NM, NJ, NY, OH, PA,TX, VA, VT, WA and WI.
- Lot, serial or model codes
- Catalog No: 6541-5-723 Lot Nos.: AF1C00, AF1C00A, AF1C00S1, AF1C11, AF1C11A, AF1C11AR, AF1C11AR1, AF1C11J, AF1C11JR, AF1C11M, AF1C11MT, AF1C11X1, AF1C12, AF1C12R, AF1E01, AF1E01D, AF1E01D1, AF1E01D1R, AF1E01M, AF1E01MR, AF1E01MR1, AF1E01MX1, AF1E01MX2, AF1E01MX2R, AF1E01MX3, AF1E01R, AF1E01R1, AF1E07, AF1E08, AF1K14, AF1K14R, AF1T43, AF1W33, AF1W34, AF1W36, AF2C11, AF2C11A, AF2C11G, AF2C11R, AF2C12, AF2C12A, AF2C12A1, AF2C12R, AF2C53, AF2C53A, AF2C58, AF2C58M, AF2C58R, AF2S08, AF3L05, AF3L05P, AF3S11, AF3T02, AF3T02J, AF3V02, AF3V02J, AF3V09, AF3V09A, AF3W04, AF3W04D, AF3W29, AF3W29D, AF3W29X1, AF4C07, AF4C07M, AF4C07P, AF4C07X1, AF5A01, AF5A02, AF5A03, AF5K00, AF5S00, AF5S00T, AF5V01, AF5V01H, AF6E03, AF6E03N, AF6E03T, AF6E03TE, AF6E04, AF6E04J, AF6E04N, AF6K00, AF6K00A, AF6K00D, AF6K00L, AF6K05, AF6K05D, AF6K05H, AF6K05X, AF7A00, AF7A00L, AF7C01. AF7C01E, AF7E00, AF7H01, AF8S00, AF8V03, AF8V03A, AF8W00, AF9E00, AF9L00, AF9L00A, AF9N00, AF9T00, AF9T00E, AFZV06, AFZV18, AFZV18A, AFZV18R, AFZW21, AFZW21A, AFZW22, AFZW23, AFZW23A, AFZW23R, AFZW23R1, AFZW23R2, AFZW23W, AFZW23X1 and AFZW23X1R
- 510(k) numbers
- ["K123486"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28