Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics, Inc.: VITROS Immunodiagnostic Products Anti-HBc IgM Controls, Product Code 6801427 For use in monitoring the performance of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems

Recall number
Z-0247-2022
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Immunodiagnostic Products Anti-HBc IgM Controls, Product Code 6801427 For use in monitoring the performance of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems
Status
Terminated
Initiated
2021-10-15
Posted
not on file
Terminated
2023-12-18
Reason
Ortho Clinical Diagnostics confirmed a complaint that some VITROS Immunodiagnostic Products Anti-HBc IgM Controls Lot 1690 Controls Values booklet contained information for the incorrect product. While the accuracy of patient test results is not affected by this issue, there is a potential for delayed results due to the troubleshooting of Quality Control results.
Root cause
Under Investigation by firm
Action
On October 15, 2021, the firm issued "Important Product Correction Notification" letters to affected customers. Customers were informed that up to 8 booklets in the affected lot contained the range of means information for VITROS Immunodiagnostic Products Anti-HBc Controls instead of the control values for VITROS Anti-HBc IgM. Customers were provided with a copy of the correct Controls Values booklet. Customers were asked to do the following: - Complete and return the included Confirmation of Receipt - Response Required customer response form by email to: ra-ocdus-confirmad@orthoclinicaldiagnostics.com or fax to: 1-888-557-3759 or 1-585-453-4110 - Retain the included Controls Values booklet to use in place of any recalled booklets - Forward the notification if the product was distributed outside your facility If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Quantity
191 units
Distribution
US (Domestic) distribution to states of: AZ, CA, CO, CT, FL, GA, LA, MA, MO, NJ, NY, OK, SC, SD, TX, UT, WA, WI, and WV. OUS (Foreign) distribution to Colombia.
Lot, serial or model codes
Lot 1690 UDI 10758750001200
510(k) numbers
["K032993"]
PMA numbers
not on file
Product code
JJX
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register