Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular Us Sales, Llc: Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.

Recall number
Z-0247-2016
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Product
Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.
Status
Terminated
Initiated
2015-08-17
Posted
2015-11-09
Terminated
2016-10-19
Reason
Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.
Root cause
Under Investigation by firm
Action
MAQUET sent an "Urgent Product Recall Medical Device Removal" notice and "Medical Device Removal Response Form" dated August 17, 2015 via UPS to affected customers. The notification identified the issue and the affected product. The notification instructed customers to read the notification and disseminate the information provided to all the users of the device; examine their inventory to determine if they have the Adjustable Positioning Elements; if so, remove the affected products, quarantine them and keep them in a secure location for shipment back to Maquet; and fill out and sign the accompany response form and send via email or fax to Maquet. If the customers have any questions regarding this field action, they were instructed to please contact their local MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383 (Press option 2, then followed by option 3) Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. EDT.
Quantity
United States (3 sets); Rest of World (47 sets)
Distribution
Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and United Arab Emirates.
Lot, serial or model codes
Part Nos 1007.28A0 (Adjustable Positioning Elements and 1007.24F0 (Universal Frame with Adjustable Positioning Elements)
510(k) numbers
not on file
PMA numbers
not on file
Product code
FWZ
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register