Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems: SKYLight Gamma Camera System; SKYLIGHT 8FT, 3/8" SPECT Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight Gamma Camera System is intended to product images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

Recall number
Z-0247-2012
Recalling firm
Philips Medical Systems
Product
SKYLight Gamma Camera System; SKYLIGHT 8FT, 3/8" SPECT Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight Gamma Camera System is intended to product images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
Status
Terminated
Initiated
2011-09-27
Posted
2011-11-19
Terminated
2012-12-21
Reason
A mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
Root cause
Device Design
Action
Philips Medical Systems sent a "Customer Information" letter dated September 27, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter lists additional precautions the customers should use while operating the device. Philips will notify customers for replacing the affected product through an Action for Performance-Proactive Field Change Order to resolve the issue. Contact Philips Customer Care Solutions at 1-800-722-9377, option 5, for questions concerning this notice.
Quantity
658 - all units in recall
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico.
Lot, serial or model codes
Model number: 2161-3000A
510(k) numbers
["K000908"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register