Skip to the content

FDA Medical Device Recalls (openFDA)

Customed, Inc: Surgeon Gown Poly Reinforced Lgc, Catalog number 900-522. Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Recall number
Z-0246-2015
Recalling firm
Customed, Inc
Product
Surgeon Gown Poly Reinforced Lgc, Catalog number 900-522. Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Status
Terminated
Initiated
2014-10-08
Posted
2014-12-16
Terminated
2017-08-14
Reason
Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.
Root cause
Packaging
Action
Customed initiated a recall on October 8th, 2014 of sterile convenience surgical Packs-trays, bags (known as Convenience Packs) and new lots of products added to the initial June 3, 2014 recall of 233 sterile surgical packs. The products are being recalled because of significant compromised sterility, not limited to potentially damaged packaging. Customers were instructed to quarantine and cease use of the affected product identified. Customers are to return the response form included with the letter, to the contact information provided. Upon receipt, the firm will contact customers with further instructions to obtain credit, replacement product, and for return of the recalled product or any other disposition action. If customers have any questions, they should contact the Customed Recall Coordinator at 407-850-5558 extensions 8001 or 8005, or via e-mail: pablo.diaz@customedhealing.com.
Quantity
86912
Distribution
US Distribution to Florida, New York and Puerto Rico..
Lot, serial or model codes
Lot numbers: 111103265 111113126 112010183 112020505 112030804 112041056 112051993 112072804 112093631 112104361 112104533 112124897 113026196 113036804 113057491 113067873 113078333 113088873 113099438 113109852 131110265 131210654 140111126 140211540 140312173 140312384 140513370 140714295 140814813 140814957 140915481
510(k) numbers
not on file
PMA numbers
not on file
Product code
FYA
Firm FEI
2648727
Firm city
Fajardo
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register