Laboratoires Bodycad, Inc.: Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.

Recall number
Z-0244-2022
Recalling firm
Laboratoires Bodycad, Inc.
Product
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Status
Terminated
Initiated
2021-07-02
Posted
not on file
Terminated
2023-01-04
Reason
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Root cause
Nonconforming Material/Component
Action
The recalling firm issued an Advisory Notice Information dated 2020-06-30 on 7/2/2021 via email which explained the reason for recall and that they would be contacted as soon as possible to proceed with the exchange. On 9/24/2021, a Returned Goods Authorization Form was sent to the consignees so the adapter could be exchanged.
Quantity
10 in the U.S.
Distribution
Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Lot, serial or model codes
Torque Limiting Adaptor - Lot number SB000430.  Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3012086398
Firm city
Quebec
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register