Skip to the content

FDA Medical Device Recalls (openFDA)

Applied Biotech Inc: Clearview hCG Combo with Reference Line, Wampole Laboratories, ABI List No/Product No. 6028KCV; 30 Individually pouched test cassettes with disposable pipettes, 1 Directional Insert; Inverness Medical Professional Diagnostics Group; Marketed by Wampole Laboratories, 2 Research Way, Princeton NJ 08540 USA

Recall number
Z-0244-2008
Recalling firm
Applied Biotech Inc
Product
Clearview hCG Combo with Reference Line, Wampole Laboratories, ABI List No/Product No. 6028KCV; 30 Individually pouched test cassettes with disposable pipettes, 1 Directional Insert; Inverness Medical Professional Diagnostics Group; Marketed by Wampole Laboratories, 2 Research Way, Princeton NJ 08540 USA
Status
Terminated
Initiated
2006-09-28
Posted
2007-11-21
Terminated
2012-08-03
Reason
Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product.
Root cause
Process design
Action
A Product Recall Notification dated 9/28/2006, was provided via fax, email or direct mail to all primary consignees. Customers were instructed to block all stock of affected lots in their warehouses, contact their primary consignees and advise them to discontinue use immediately and dispose of any unused materials; using a fax back form to communicate to the recalling firm the quantity blocked and disposed of so that arrangements could be made for providing replacement product
Quantity
400 Kits (1,200 tests)
Distribution
Worldwide: USA and Canada
Lot, serial or model codes
Model Number: 6028KCV, ABI Lot/Unit Code: 88982
510(k) numbers
not on file
PMA numbers
not on file
Product code
JHI
Firm FEI
1000125390
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register