FDA Medical Device Recalls (openFDA)
Texas Medical Technology Inc.: Disposable Surgical Level 3 Gown, Size/Product Code: Large/GWS-03-L-B, XLarge/GWS-03-XL-B, XXLarge/GWS-03-XXL-B
- Recall number
- Z-0242-2023
- Recalling firm
- Texas Medical Technology Inc.
- Product
- Disposable Surgical Level 3 Gown, Size/Product Code: Large/GWS-03-L-B, XLarge/GWS-03-XL-B, XXLarge/GWS-03-XXL-B
- Status
- Open, Classified
- Initiated
- 2022-09-01
- Posted
- 2022-11-16
- Terminated
- not on file
- Reason
- Surgical gowns recalled due to gown fabric failing to pass the hydrostatic pressure test so the fabric does not provide the protection needed in a level 3 surgical gown, i.e., blood and other fluid may penetrate the fabric.
- Root cause
- Nonconforming Material/Component
- Action
- On 9/1/22, a recall notice was sent to a distributor who was asked to forward the recall notice to their customers. Customers are asked to return the gowns to the distributor and to complete a self-address questionnaire to the recalling firm. The following contact information was provided: 346-250-2655, Monday through Friday 8 AM to 5 PM Central Time, sales@texasmedicaltechnology.com On 11/9/22, a follow-up recall notice was sent asking customers to destroy affected gowns.
- Quantity
- 100,000
- Distribution
- US Nationwide distribution in the state of FL.
- Lot, serial or model codes
- Product Code/UDI-DI/Lots: GWS-03-L-B/00850022960338/1300001, 1300002, 1300003, 1300004; GWS-03-XL-B/00850022960345/1400001, 1400002, 1400003, 1400004; GWS-03-XXL-B/00850022960352/1500001, 1500002, 1500003, 1500004
- 510(k) numbers
- ["K120192"]
- PMA numbers
- not on file
- Product code
- FYA
- Firm FEI
- 3016808907
- Firm city
- Sugar Land
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28