Skip to the content

FDA Medical Device Recalls (openFDA)

Texas Medical Technology Inc.: Disposable Surgical Level 3 Gown, Size/Product Code: Large/GWS-03-L-B, XLarge/GWS-03-XL-B, XXLarge/GWS-03-XXL-B

Recall number
Z-0242-2023
Recalling firm
Texas Medical Technology Inc.
Product
Disposable Surgical Level 3 Gown, Size/Product Code: Large/GWS-03-L-B, XLarge/GWS-03-XL-B, XXLarge/GWS-03-XXL-B
Status
Open, Classified
Initiated
2022-09-01
Posted
2022-11-16
Terminated
not on file
Reason
Surgical gowns recalled due to gown fabric failing to pass the hydrostatic pressure test so the fabric does not provide the protection needed in a level 3 surgical gown, i.e., blood and other fluid may penetrate the fabric.
Root cause
Nonconforming Material/Component
Action
On 9/1/22, a recall notice was sent to a distributor who was asked to forward the recall notice to their customers. Customers are asked to return the gowns to the distributor and to complete a self-address questionnaire to the recalling firm. The following contact information was provided: 346-250-2655, Monday through Friday 8 AM to 5 PM Central Time, sales@texasmedicaltechnology.com On 11/9/22, a follow-up recall notice was sent asking customers to destroy affected gowns.
Quantity
100,000
Distribution
US Nationwide distribution in the state of FL.
Lot, serial or model codes
Product Code/UDI-DI/Lots: GWS-03-L-B/00850022960338/1300001, 1300002, 1300003, 1300004; GWS-03-XL-B/00850022960345/1400001, 1400002, 1400003, 1400004; GWS-03-XXL-B/00850022960352/1500001, 1500002, 1500003, 1500004
510(k) numbers
["K120192"]
PMA numbers
not on file
Product code
FYA
Firm FEI
3016808907
Firm city
Sugar Land
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register