GE Medical Systems, LLC: SIGNA Architect system Product Usage: The Signa Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Recall number
Z-0236-2019
Recalling firm
GE Medical Systems, LLC
Product
SIGNA Architect system Product Usage: The Signa Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Status
Terminated
Initiated
2018-10-03
Posted
not on file
Terminated
2022-05-27
Reason
A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Root cause
Device Design
Action
On about 10/03/2018, GE Healthcare sent URGENT MEDICAL DEVICE CORRECTION letters to their consignees informing them of recall and providing safety instructions. Consignees were instructed that they can continue to use the system by following the existing Operator Manual Safety Information for proper patient padding and observation during scanning. The safety information contained in the Operator's Manual includes: 1. WARNING: Place appropriate non-conductive padding between the patient and the bore wherever a portion of the body may come into contact with the magnet opening. 2. CAUTION: Continuous patient observation and contact are required in all modes of operation. 3. CAUTION: Extra attention should be utilized when scanning patients who are unconscious, sedated, or may have loss of feeling in any body part or are physically or mentally impaired and unable to alert the personnel. If a brown discoloration is seen on the inner surface of the patient bore, or if a burning smell is noticed, please contact their local Service Representative. GE Healthcare will contact consignees to arrange for the correction to the devices. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
Quantity
4 units
Distribution
Worldwide - US Nationwide Distribution in the state of Utah. International distribution to France, Republic of Korea, and Sweden.
Lot, serial or model codes
Model number: 5388807-100, Serial number (System ID number ): U30RF0251B (801588MR3), U30RF0258B (M4040093), U30RF0275B (850270504), U30RF0268B (A020MR05)
510(k) numbers
["K160618"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2183553
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register