Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems, Inc.: Philips Brilliance iCT Computed Tomography X-Ray System Product Usage: The Brilliance iCT/Brilliance iCT SP is a Computed Tomography X-Ray System intended to produce crosssectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. (d) A description of the defect in the

Recall number
Z-0236-2015
Recalling firm
Philips Medical Systems, Inc.
Product
Philips Brilliance iCT Computed Tomography X-Ray System Product Usage: The Brilliance iCT/Brilliance iCT SP is a Computed Tomography X-Ray System intended to produce crosssectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. (d) A description of the defect in the
Status
Terminated
Initiated
2014-07-31
Posted
2014-12-16
Terminated
2017-02-16
Reason
It was discovered that a software defect may result in the scanner not terminating the CT scan at the intended location.
Root cause
Radiation Control for Health and Safety Act
Action
Philips Healthcare is directly notifying affected users via letter. In addition, Philips Healthcare has performed the software correction for forward production and plans a Field Correction to fix the defect in the field. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office.
Quantity
16 systems
Distribution
US Distribution
Lot, serial or model codes
Software version 4.1.
510(k) numbers
["K060937"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28