Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Sustainability Solutions: SSS Reprocessed Zimmer, Synthes, Stryker and Linvatec Drill Bits and Blades Surgical drill bits, blades, burs, chisels, rasps, taps and reamers are specially designed surgical accessories to help the surgeon cut and drill bone and bony structures. Reprocessed surgical accessories are indicated for use in patients requiring surgery involving bone and bony structures.

Recall number
Z-0235-2016
Recalling firm
Stryker Sustainability Solutions
Product
SSS Reprocessed Zimmer, Synthes, Stryker and Linvatec Drill Bits and Blades Surgical drill bits, blades, burs, chisels, rasps, taps and reamers are specially designed surgical accessories to help the surgeon cut and drill bone and bony structures. Reprocessed surgical accessories are indicated for use in patients requiring surgery involving bone and bony structures.
Status
Terminated
Initiated
2015-10-14
Posted
not on file
Terminated
2016-05-06
Reason
It was reported that packaging was not properly sealed.
Root cause
Employee error
Action
Stryker sent an Urgent Medical Device Recall letter dated October 14, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm's letter states that the use of the affected lots of the recalled product should be discontinued. Other bit and/or blade lots reprocessed by Stryker Sustainability Solutions should be considered as alternatives. A Recall Effectiveness Check Form including the quantity of affected devices shipped is enclosed. The letter states to complete the Recall Effectiveness Check Form and indicate if any devices from the affected lots remain in inventory. The form needs to be completed even if no affected product is found. Return the completed and signed form to the local Stryker Sustainability Sales Representative or email to to SSSPFA@stryker.com If the form indicates that affected devices remain in inventory, a prepaid shipping label will be issued for the return of the product. Customers will receive a credit for all affected devices returned. For further questions, please call (888) 888-3433.
Quantity
1202
Distribution
US ( nationwide) Distribution to the states of : MN, OH, CA, IN, WI, FL, GA, MO, IA, MT, LA, KS, TX, TN, WY, MD, NJ, WA, PA, SC, NY, MI, AT, HI, MN, AR, AZ, IL, IA, KY, MS, UT, NC, VA, CO, ME and ID., Internationally to Canada.
Lot, serial or model codes
Item No/Model No: 2108-109, 4125-127-090, 2108-110, 2108-115, 2108-120, 2108-151-001, 2108-152, 2108-158, 2108-182, 2108-218, 310.25, 4118-127-090, 4125-147-090, 4806-110-25, 5071-553, 6118-127-90, 6125-127-090  Lot Code: 3743743, 3750611, 3750630, 3753894, 3761982, 3775827, 3779058, 3779129, 3782549, 3782554, 3782565, 3785624, 3785630, 3785676, 3789092, 3789102, 3794000, 3794085, 3796706, 3796799, 3799827, 3802103
510(k) numbers
not on file
PMA numbers
not on file
Product code
GFA
Firm FEI
3009350690
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register