Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens SimView NT and SimView NT with IDI simulation system components: Product Usage: The intended use of the Radiation Therapy Simulation system is used for simulation of treatment planning prior to the therapeutic treatment of cancer.

Recall number
Z-0235-2015
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens SimView NT and SimView NT with IDI simulation system components: Product Usage: The intended use of the Radiation Therapy Simulation system is used for simulation of treatment planning prior to the therapeutic treatment of cancer.
Status
Terminated
Initiated
2014-10-10
Posted
2014-11-20
Terminated
2017-06-29
Reason
Siemens became aware of a very rare occurrence where an intended movement of the SimView NT simulation system components could not be stopped with usual hand control commands or by use of the collision detection functionality. It is possible that simulation system components could collide with patient and/or system equipment.
Root cause
Device Design
Action
Customers were initially notified about the issue by letter on January 7, 2014 (this action was reported as Z-0786-2014 which has been terminated). Siemens promised to notify them when a fix became available. Customer letter, dated November 7, 2014, was sent to notify end users that a fix has become available and Siemens would contact them to schedule an appointment for the correction.
Quantity
6
Distribution
Worldwide Distribution - US Nationwide in the states of PA, KY, LA, MA, MI, and CA; and countries of: Argentina, Belarus, Bulgaria, Costa Rico, P.R. China, Curacao, Egypt, Repulic Korea, Germany, Greece, Hungary, Italy, Iraq, Jordan, Malaysia, Palestanian Ter, Poland, Thailand, Turkey, and Vietnam.
Lot, serial or model codes
Model numbers: 5496992, 5497016   with Serial numbers: 6035, 6080, 6029, 6032, 6019, 6001
510(k) numbers
["K953233"]
PMA numbers
not on file
Product code
KPQ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register