Skip to the content

FDA Medical Device Recalls (openFDA)

Pajunk GmbH: SPROTTE Lumbar with Introducer, labeled as the following: 1. Sprotte 22G x 90mm w/ introducer and wings (25/box); 2. Sprotte 22G x 120mm w/ introducer and wings (25/box); 3. Sprotte 19G x 90mm w/ introducer and wings (25/box]); 4. Sprotte 22G x 103mm w/ introducer and wings (25/box); 5. Sprotte 21G x 90mm w/ introducer and wings (25/box); 6. Sprotte 21G x 103mm w/ introducer and wings (25/box); 7. Sprotte 20G x 103mm w/ introducer and wings (25/box); 8. Sprotte 20G x 90mm w/ introducer and wings (25/box) Product Usage: The SPROTTE lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography procedures. The device is intended for adult and pediatric patients.

Recall number
Z-0234-2020
Recalling firm
Pajunk GmbH
Product
SPROTTE Lumbar with Introducer, labeled as the following: 1. Sprotte 22G x 90mm w/ introducer and wings (25/box); 2. Sprotte 22G x 120mm w/ introducer and wings (25/box); 3. Sprotte 19G x 90mm w/ introducer and wings (25/box]); 4. Sprotte 22G x 103mm w/ introducer and wings (25/box); 5. Sprotte 21G x 90mm w/ introducer and wings (25/box); 6. Sprotte 21G x 103mm w/ introducer and wings (25/box); 7. Sprotte 20G x 103mm w/ introducer and wings (25/box); 8. Sprotte 20G x 90mm w/ introducer and wings (25/box) Product Usage: The SPROTTE lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography procedures. The device is intended for adult and pediatric patients.
Status
Open, Classified
Initiated
2019-08-30
Posted
not on file
Terminated
not on file
Reason
Problem with packaging sealing process which affects sterility.
Root cause
Process control
Action
Pajunk GmbH Medizintechnologie notified distributor and end customers on about 08/30/2019, via "Urgent Product Recall Notice" letters. Instructions included to immediately identify any affected products in inventory and quarantine them, cease use of affected products, and complete and return the Attachment B indicating receipt of the notice. Additionally end customers and distributors were instructed to provide the recall notice to all responsible personnel within the organization or if products were further distributed. Questions or concerns can be directed to Laura Roberts, Manager of Office of Operations and Client Relations, email Laura.roberts@pejunk-usa.com, phone (470) 448-4075.
Quantity
40,475 devices
Distribution
Worldwide distributions - US Nationwide including the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, RI, SC, TN, TX, WI, WY. Countries of Germany, Ellada / Greece, Canada, Italy, Lietuva / Lithuania/ Litauen, Australia, Luxembourg, Austria, Nederland / Netherlands, Belgium, Polska / Poland, Switzerland, Sverige / Sweden, Czech Republic, Slovenija / Slovenia, Danmark / Denmark, Kroatien/ Croatia, Espa¿a / Spain, Aserbeidschan/ Azerbaidzan, Suomi / Finland, Brunei, France, Russland/ Russia, Great Britain, Oman, Israel.
Lot, serial or model codes
Lot Numbers 1240 through 1313. Item numbers:   1. 321151-30C;   2. 331151-30C;   3. 321151-31C;   4. 341151-30C;   5. 321151-31A;   6. 341151-31A;   7. 341151-31B;   8. 331151-31B
510(k) numbers
["K911260"]
PMA numbers
not on file
Product code
BSP
Firm FEI
3002807847
Firm city
Geisingen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register