Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta Inc: Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.

Recall number
Z-0234-2018
Recalling firm
Elekta Inc
Product
Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.
Status
Terminated
Initiated
2017-09-01
Posted
not on file
Terminated
2019-11-05
Reason
Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.
Root cause
Nonconforming Material/Component
Action
Elekta sent an Urgent Important Field Safety Notification letter dated September, 2017. The letter identified affected product, problem and actions to be taken. Customers were instructed that a copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. The customers were asked to complete the verification tests and return the Important Field Safety Notice Acknowledgment form to their local Elekta representative as soon as possible but no later than 30 days from receipt. For questions contact your local Elekta Care Support Center.
Quantity
182
Distribution
Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM
Lot, serial or model codes
Version 1.7.3 and higher
510(k) numbers
["K080601"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
1937649
Firm city
Maryland Heights
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register