FDA Medical Device Recalls (openFDA)
Elekta Inc: Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.
- Recall number
- Z-0234-2018
- Recalling firm
- Elekta Inc
- Product
- Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.
- Status
- Terminated
- Initiated
- 2017-09-01
- Posted
- not on file
- Terminated
- 2019-11-05
- Reason
- Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.
- Root cause
- Nonconforming Material/Component
- Action
- Elekta sent an Urgent Important Field Safety Notification letter dated September, 2017. The letter identified affected product, problem and actions to be taken. Customers were instructed that a copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. The customers were asked to complete the verification tests and return the Important Field Safety Notice Acknowledgment form to their local Elekta representative as soon as possible but no later than 30 days from receipt. For questions contact your local Elekta Care Support Center.
- Quantity
- 182
- Distribution
- Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM
- Lot, serial or model codes
- Version 1.7.3 and higher
- 510(k) numbers
- ["K080601"]
- PMA numbers
- not on file
- Product code
- MUJ
- Firm FEI
- 1937649
- Firm city
- Maryland Heights
- Firm state
- MO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28