Skip to the content

FDA Medical Device Recalls (openFDA)

Miramar Labs, Inc.: miraDry System consisting of the MD4000-MC Console, the MD4000-HP miraDry Handpiece, and the MD4000-BT miraDry bioTip and accessories. General and Plastic Surgery: The miraDry System is intended for the treatment of primary axillary hyperhidrosis and was recently cleared for the permanent removal of underarm hair.

Recall number
Z-0234-2016
Recalling firm
Miramar Labs, Inc.
Product
miraDry System consisting of the MD4000-MC Console, the MD4000-HP miraDry Handpiece, and the MD4000-BT miraDry bioTip and accessories. General and Plastic Surgery: The miraDry System is intended for the treatment of primary axillary hyperhidrosis and was recently cleared for the permanent removal of underarm hair.
Status
Terminated
Initiated
2015-10-09
Posted
2015-11-06
Terminated
2015-12-23
Reason
Update to labeling for the miraDry Console to remind users to only use approved lubricants in a thin layer to prevent potential burns and blisters.
Root cause
Labeling design
Action
Miramar Labs sent Medical Device Correction letters to all users on October 9, 2015 to inform them of the labeling clarification by UAPS First Class Certified mail or courier such as Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that the firm added a new warning to their labeling. The firm included an updated copy of the MD4000-MC User Manual with the new warning included and asked that customers destroy their older copy. Customers with questions were instructed to call Customer SService at 1-855-647-2379 or email info@miraDry.com
Quantity
277 units
Distribution
Nationwide Distribution
Lot, serial or model codes
MN-MD4000-MC, all units
510(k) numbers
["K103014","K131162","K150419"]
PMA numbers
not on file
Product code
OUB
Firm FEI
3008082710
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register