CPM Medical Consultants, LLC.: Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS200326, OS328026, OS328028C, OS216150, OS218150, OS292110, OS292120, OS292160, OS292200, OS200516, OS200518, OS200520, OS200522, OS200616, OS200618, OS200620, OS200622, OS421300, OS421302, OS421304, OS421306, OS421308, OS421208, OS421210, OS421212, OS421000L, OS421000R, OS421002L, OS421002R, OS421004L, OS421004R, OS421006L, OS421006R, OS42260LL, OS42260LR, OS42260ML, OS42260MR, OS42260SL, OS42260SR, OS42260XSL, OS42260XSR, OS422800, OS422803, OS422804, OS422805, OS422807, OS422825, OS422400, OS422403, OS422404, OS422405, OS422406, OS422407, OS421406, OS421408, OS421414, OS422702, OS422703, OS422712, OS422714, OS422716, OS421512, OS421514, OS421516, OS421112, OS421116, OS421120, OS421124, OS421130, OS42150L, OS42150M, OS42150S, OS42150XS, OS326920-T, OS326922-T, OS326924-T, OS326925, OS326925-T, OS326926-T, OS326928, OS326928-T, OS326930, OS326930-T, OS326932, OS326932-T, OS326934, OS326934-T, OS326936, OS326936-T, OS326938, OS326938-T, OS326940, OS326940-T, OS326945, OS326945-T, OS326950, OS326950-T, OS326955, OS326955-T, OS326960, OS326960-T, OS327920, OS327922, OS327924, OS327926, OS327928, OS327930, OS327932, OS327934, OS327936, OS327938, OS327940, OS327945, OS327950, OS327955, OS327960, OS421710, OS421712, OS421714, OS421716, OS421718, OS421720, OS421722, OS421724, OS421726, OS421728, OS421730, OS421735, OS421740, OS421745, OS421750, OS422508, OS422510, OS422512, OS422514, OS422516, OS422518, OS422520, OS422522, OS422524, OS422526, OS422528, OS422530, OS421814, OS421816, OS421818, OS421820, OS421822, OS421824, OS421826, OS421828, OS421830, OS421835, OS421840, OS421845, OS421850, OS422608, OS422610, OS422612, OS422614, OS422616, OS422618, OS422620, OS422622, OS422624, OS422626, OS422628, OS422630, R327924, R327928, R327932, R327936, R327940, R327945, R327950, R421000L, R421000R, R421710, R421712, R421714, R421716, R421718, R421720, R421722, R421724, R421726, R421728, R421730, R421735, R421740, R421814, R421816, R421818, R421820, R421822, R421824, R421826, R421828, R421830, R421835, R421840, R421845, R421850, R42260ML, R42260MR, R42260SL, R42260SR, R42260XSL, R42260XSR

Recall number
Z-0230-2026
Recalling firm
CPM Medical Consultants, LLC.
Product
Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS200326, OS328026, OS328028C, OS216150, OS218150, OS292110, OS292120, OS292160, OS292200, OS200516, OS200518, OS200520, OS200522, OS200616, OS200618, OS200620, OS200622, OS421300, OS421302, OS421304, OS421306, OS421308, OS421208, OS421210, OS421212, OS421000L, OS421000R, OS421002L, OS421002R, OS421004L, OS421004R, OS421006L, OS421006R, OS42260LL, OS42260LR, OS42260ML, OS42260MR, OS42260SL, OS42260SR, OS42260XSL, OS42260XSR, OS422800, OS422803, OS422804, OS422805, OS422807, OS422825, OS422400, OS422403, OS422404, OS422405, OS422406, OS422407, OS421406, OS421408, OS421414, OS422702, OS422703, OS422712, OS422714, OS422716, OS421512, OS421514, OS421516, OS421112, OS421116, OS421120, OS421124, OS421130, OS42150L, OS42150M, OS42150S, OS42150XS, OS326920-T, OS326922-T, OS326924-T, OS326925, OS326925-T, OS326926-T, OS326928, OS326928-T, OS326930, OS326930-T, OS326932, OS326932-T, OS326934, OS326934-T, OS326936, OS326936-T, OS326938, OS326938-T, OS326940, OS326940-T, OS326945, OS326945-T, OS326950, OS326950-T, OS326955, OS326955-T, OS326960, OS326960-T, OS327920, OS327922, OS327924, OS327926, OS327928, OS327930, OS327932, OS327934, OS327936, OS327938, OS327940, OS327945, OS327950, OS327955, OS327960, OS421710, OS421712, OS421714, OS421716, OS421718, OS421720, OS421722, OS421724, OS421726, OS421728, OS421730, OS421735, OS421740, OS421745, OS421750, OS422508, OS422510, OS422512, OS422514, OS422516, OS422518, OS422520, OS422522, OS422524, OS422526, OS422528, OS422530, OS421814, OS421816, OS421818, OS421820, OS421822, OS421824, OS421826, OS421828, OS421830, OS421835, OS421840, OS421845, OS421850, OS422608, OS422610, OS422612, OS422614, OS422616, OS422618, OS422620, OS422622, OS422624, OS422626, OS422628, OS422630, R327924, R327928, R327932, R327936, R327940, R327945, R327950, R421000L, R421000R, R421710, R421712, R421714, R421716, R421718, R421720, R421722, R421724, R421726, R421728, R421730, R421735, R421740, R421814, R421816, R421818, R421820, R421822, R421824, R421826, R421828, R421830, R421835, R421840, R421845, R421850, R42260ML, R42260MR, R42260SL, R42260SR, R42260XSL, R42260XSR
Status
Open, Classified
Initiated
2025-09-17
Posted
2025-10-21
Terminated
not on file
Reason
Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.
Root cause
Labeling design
Action
On 9/17/2025, correction notices were mailed to customers and distributors who were informed of the following: - The only sterilization requirement now shown on the new Rev D IFU is 4 mins exposure time at 132-Degrees C (270-Degrees F) and 20-30mins drying time. - All affected versions of the IFU should be immediately discarded. - You can access the e-IFU at: https://fusemedical.com/wp-content/uploads/2025/09/Fuse-ULTRA-Foot-Plating-System-IFU-Rev-D-9.2025.pdf - Please pass on this Medical Device Correction to all users of the products and all other people who need to be aware within your organization. - If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice, response form, and updated IFU. - Complete and return the response form via email to FMohammad@FuseMedical.com If you have any questions, contact the firm at info@fusemedical.com
Quantity
3317
Distribution
US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.
Lot, serial or model codes
All versions of IFU DC-0495 prior to Rev D. All Lots Up to Date 2025/09/10. UDI-DI: B565OS200020L, B565OS200025L, B565OS200120L, B565OS200125, B565OS200125L, B565OS200135, B565OS200324, B565OS200325, B565OS200326, B565OS200400, B565OS200401, B565OS200516, B565OS200518, B565OS200520, B565OS200522, B565OS200616, B565OS200618, B565OS200620, B565OS200622, B565OS216150, B565OS218150, B565OS292110, B565OS292120, B565OS292160, B565OS292200, B565OS326920T, B565OS326922T, B565OS326924T, B565OS326925, B565OS326925T, B565OS326926T, B565OS326928, B565OS326928T, B565OS326930, B565OS326930T, B565OS326932, B565OS326932T, B565OS326934, B565OS326934T, B565OS326936, B565OS326936T, B565OS326938, B565OS326938T, B565OS326940, B565OS326940T, B565OS326945, B565OS326945T, B565OS326950, B565OS326950T, B565OS326955, B565OS326955T, B565OS326960, B565OS326960T, B565OS327920, B565OS327922,B565OS327924, B565OS327926, B565OS327928, B565OS327930, B565OS327932, B565OS327934, B565OS327936, B565OS327938, B565OS327940, B565OS327945, B565OS327950, B565OS327955, B565OS327960, B565OS328017, B565OS328026, B565OS328028C, B565OS421000L, B565OS421000R, B565OS421002L, B565OS421002R, B565OS421004L, B565OS421004R, B565OS421006L, B565OS421006R, B565OS421112, B565OS421116, B565OS421120, B565OS421124, B565OS421130, B565OS421208, B565OS421210, B565OS421212, B565OS421300, B565OS421302, B565OS421304, B565OS421306, B565OS421308, B565OS421406, B565OS421408, B565OS421414, B565OS42150L, B565OS42150M, B565OS42150S, B565OS42150XS, B565OS421512, B565OS421514, B565OS421516, B565OS421710, B565OS421712, B565OS421714, B565OS421716, B565OS421718, B565OS421720, B565OS421722, B565OS421724, B565OS421726, B565OS421728, B565OS421730, B565OS421735, B565OS421740, B565OS421745, B565OS421750, B565OS421814, B565OS421816, B565OS421818, B565OS421820, B565OS421822, B565OS421824, B565OS421826, B565OS421828, B565OS421830, B565OS421835, B565OS421840, B565OS421845, B565OS421850, B565OS422400, B565OS422403, B565OS422404, B565OS422405, B565OS422406, B565OS422407, B565OS422508, B565OS422510, B565OS422512, B565OS422514, B565OS422516, B565OS422518, B565OS422520, B565OS422522, B565OS422524, B565OS422526, B565OS422528, B565OS422530, B565OS422608, B565OS42260LL, B565OS42260LR, B565OS42260ML, B565OS42260MR, B565OS42260SL, B565OS42260SR, B565OS42260XSL, B565OS42260XSR, B565OS422610, B565OS422612, B565OS422614, B565OS422616, B565OS422618, B565OS422620, B565OS422622, B565OS422624, B565OS422626, B565OS422628, B565OS422630, B565OS422702, B565OS422703, B565OS422712, B565OS422714, B565OS422716, B565OS42280, B565OS422803, B565OS422804, B565OS422805, B565OS422807, B565OS422825, B565R327924, B565R327928, B565R327932, B565R327936, B565R327940, B565R327945, B565R327950, B565R421000L, B565R421000R, B565R421710, B565R421712, B565R421714, B565R421716, B565R421718, B565R421720, B565R421722, B565R421724, B565R421726, B565R421728, B565R421730, B565R421735, B565R421740, B565R421814, B565R421816, B565R421818, B565R421820, B565R421822, B565R421824, B565R421826, B565R421828, B565R421830, B565R421835, B565R421840, B565R421845, B565R421850, B565R42260M, B565R42260MR, B565R42260SL, B565R42260SR, B565R42260XSL, B565R42260XSR
510(k) numbers
["K141784"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3007334927
Firm city
Farmers Branch
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register