FDA Medical Device Recalls (openFDA)
Boston Scientific Neuromodulation Corporation: Superion Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm
- Recall number
- Z-0230-2023
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Product
- Superion Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm
- Status
- Open, Classified
- Initiated
- 2022-10-12
- Posted
- 2022-11-14
- Terminated
- not on file
- Reason
- Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.
- Root cause
- Process change control
- Action
- On 10/12/2022, the firm, Boston Scientific, sent an "Urgent: Medical Device Recall" letter via FedEx to customers. The letter described the product, problem and actions to be taken. The Customers are instructed to: -Review their facility's device inventory: Segregate and return any unused Superion IDS devices from the affected subset to Boston Scientific. Complete and return account-specific Reply Verification Tracking Form via Email: BSCFieldActionCenter@bsci.com or Fax to: Field Action Center at 1-763-415-7708. -Patient follow-up: Discuss this advisory with their patient(s) who are implanted with an affected device. Promptly investigate any reported pain and/or discomfort, or any suspected indication of migration of the Superion IDS implant, which may suggest a potential spindle cap weld break. If an affected device remains implanted in their patient, and successful deployment and/or device structural integrity is questioned, consider confirming via imaging. -If a spindle cap weld break is confirmed post-implant: Assess the clinical benefits versus risks associated with explanting an affected Superion IDS from their patient. Note that the Superion IDS Inserter tool would be unable to engage a non-intact device for explant. Medical judgement is required for determining the appropriate surgical method for explant in such cases. -Records: For each patient with an affected Superion IDS, append their medical record with a copy of this letter to maintain awareness of this topic for the remaining service life of the device. Should any revision/replacement occur, please inform Boston Scientific with relevant clinical information. For additional information or questions - contact the local Boston Scientific representative, or the Firm's Public Contact, Director, Global Corporate Communications at 415-730-2310 or jessica.sachariason@bsci.com.
- Quantity
- 175 devices
- Distribution
- U.S Nationwide Distribution including states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MI, MN, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, WI, WV, and WY
- Lot, serial or model codes
- Model (UPN): 101-9808 UDI-DI (GTIN) Code: 00884662000529 Lot/Batch Numbers: 26176977 Model (UPN): 101-9812 UDI-DI (GTIN) Code: 00884662000543 Lot/Batch Numbers: 26573432 26375758 26815560 26807662 26654632 26928168 26929605 26489700 Model (UPN): 101-9814 UDI-DI (GTIN) Code: 00884662000550 Lot/Batch Numbers: 26366842 26848047 26619510 26323984 26406210 26799951
- 510(k) numbers
- not on file
- PMA numbers
- ["P140004"]
- Product code
- NQO
- Firm FEI
- 3006630150
- Firm city
- Valencia
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as