Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Neuromodulation Corporation: Superion Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm

Recall number
Z-0230-2023
Recalling firm
Boston Scientific Neuromodulation Corporation
Product
Superion Indirect Decompression System (IDS) Model Numbers and Size: 101-9808 - 8mm 01-9812 - 12mm 101-9814 - 14mm
Status
Open, Classified
Initiated
2022-10-12
Posted
2022-11-14
Terminated
not on file
Reason
Due to Indirect Decompression System devices exhibiting a potential for weld breakage at the spindle cap-to-main body interface. As a result of the separation of the spindle cap from the main body it may prevent proper deployment of the device during implantation.
Root cause
Process change control
Action
On 10/12/2022, the firm, Boston Scientific, sent an "Urgent: Medical Device Recall" letter via FedEx to customers. The letter described the product, problem and actions to be taken. The Customers are instructed to: -Review their facility's device inventory: Segregate and return any unused Superion IDS devices from the affected subset to Boston Scientific. Complete and return account-specific Reply Verification Tracking Form via Email: BSCFieldActionCenter@bsci.com or Fax to: Field Action Center at 1-763-415-7708. -Patient follow-up: Discuss this advisory with their patient(s) who are implanted with an affected device. Promptly investigate any reported pain and/or discomfort, or any suspected indication of migration of the Superion IDS implant, which may suggest a potential spindle cap weld break. If an affected device remains implanted in their patient, and successful deployment and/or device structural integrity is questioned, consider confirming via imaging. -If a spindle cap weld break is confirmed post-implant: Assess the clinical benefits versus risks associated with explanting an affected Superion IDS from their patient. Note that the Superion IDS Inserter tool would be unable to engage a non-intact device for explant. Medical judgement is required for determining the appropriate surgical method for explant in such cases. -Records: For each patient with an affected Superion IDS, append their medical record with a copy of this letter to maintain awareness of this topic for the remaining service life of the device. Should any revision/replacement occur, please inform Boston Scientific with relevant clinical information. For additional information or questions - contact the local Boston Scientific representative, or the Firm's Public Contact, Director, Global Corporate Communications at 415-730-2310 or jessica.sachariason@bsci.com.
Quantity
175 devices
Distribution
U.S Nationwide Distribution including states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MI, MN, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, WI, WV, and WY
Lot, serial or model codes
Model (UPN): 101-9808 UDI-DI (GTIN) Code: 00884662000529 Lot/Batch Numbers: 26176977 Model (UPN): 101-9812 UDI-DI (GTIN) Code: 00884662000543 Lot/Batch Numbers: 26573432 26375758 26815560 26807662 26654632 26928168 26929605 26489700 Model (UPN): 101-9814 UDI-DI (GTIN) Code: 00884662000550 Lot/Batch Numbers: 26366842 26848047 26619510 26323984 26406210 26799951
510(k) numbers
not on file
PMA numbers
["P140004"]
Product code
NQO
Firm FEI
3006630150
Firm city
Valencia
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register