Skip to the content

FDA Medical Device Recalls (openFDA)

Theken Companies LLC: iNSitu Bipolar Hip System, Bipolar head implant Part ID Description 10-22-0038 Bipolar Head Assembly, Size 38mm 10-22-0040 Bipolar Head Assembly, Size 40mm 10-22-0042 Bipolar Head Assembly, Size 42mm 10-22-0044 Bipolar Head Assembly, Size 44mm 10-22-0046 Bipolar Head Assembly, Size 46mm 10-22-0048 Bipolar Head Assembly, Size 48mm 10-22-0050 Bipolar Head Assembly, Size 50mm 10-22-0052 Bipolar Head Assembly, Size 52mm 10-22-0054 Bipolar Head Assembly, Size 54mm 10-22-0056 Bipolar Head Assembly, Size 56mm 10-22-0058 Bipolar Head Assembly, Size 58mm 10-22-0060 Bipolar Head Assembly, Size 60mm

Recall number
Z-0227-2026
Recalling firm
Theken Companies LLC
Product
iNSitu Bipolar Hip System, Bipolar head implant Part ID Description 10-22-0038 Bipolar Head Assembly, Size 38mm 10-22-0040 Bipolar Head Assembly, Size 40mm 10-22-0042 Bipolar Head Assembly, Size 42mm 10-22-0044 Bipolar Head Assembly, Size 44mm 10-22-0046 Bipolar Head Assembly, Size 46mm 10-22-0048 Bipolar Head Assembly, Size 48mm 10-22-0050 Bipolar Head Assembly, Size 50mm 10-22-0052 Bipolar Head Assembly, Size 52mm 10-22-0054 Bipolar Head Assembly, Size 54mm 10-22-0056 Bipolar Head Assembly, Size 56mm 10-22-0058 Bipolar Head Assembly, Size 58mm 10-22-0060 Bipolar Head Assembly, Size 60mm
Status
Open, Classified
Initiated
2025-09-11
Posted
2025-10-21
Terminated
not on file
Reason
Downstream recall for RES 97605, BioPro femoral components. Product failed transit testing. Impacted product may have potential issues with sterile barrier.
Root cause
Package design/selection
Action
NextStep Arthropedix contacted their distributors via phone on September 11-12, 2025, to notify them of the situation. The distributors were instructed to quarantine and return all Bipolar Head implants. A formal recall letter with instructions for returning the inventory to NextStep Arthropedix was sent to customers on September 19, 2025.
Quantity
409
Distribution
US Nationwide distribution in the states of NJ, PA, and TX.
Lot, serial or model codes
10-22-0038 UDI-DI 00816127029431 Lots 2036, 2036-RP, 3495, 3495-1; 10-22-0040 UDI-DI 00816127029448 Lots 2037, 2037-RP, 3491, 3491-1; 10-22-0042 UDI-DI 00816127029455 Lot 2038-RP; 10-22-0044 UDI-DI 00816127029462 Lots 2039-RP, 3496-1; 10-22-0046 UDI-DI 00816127029479 Lots 2040, 2040-RP, 3492, 3492-1; 10-22-0048 UDI-DI 00816127029486 Lot 2041-RP; 10-22-0050 UDI-DI 00816127029493 Lots 2042, 2042-RP; 10-22-0052 UDI-DI 00816127029509 Lots 2043, 2043-RP; 10-22-0054 UDI-DI 00816127029516 Lots 2044, 2044-RP; 10-22-0056 UDI-DI 00816127029523 Lots 2045, 2045-RP; 10-22-0058 UDI-DI 00816127029530 Lots 2046, 2046-RP, 3493; 10-22-0060 UDI-DI 00816127029547 Lots 2047, 2047-RP, 3494 All lots/serial numbers
510(k) numbers
["K191297","K241875"]
PMA numbers
not on file
Product code
MEH
Firm FEI
3002498892
Firm city
Akron
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register