FDA Medical Device Recalls (openFDA)
Roche Diagnostics Corp.: CoaguChek brand PT Test Strips; U.S. Catalog Number 3116247 (48 strip pack-professional use), 3116239 (12 strip pack-patient self test), Non-U.S. Catalog Numbers 11937642190 [packaged 48 strips per box] and 11937634190 [packaged 12 strips per box].
- Recall number
- Z-0227-2007
- Recalling firm
- Roche Diagnostics Corp.
- Product
- CoaguChek brand PT Test Strips; U.S. Catalog Number 3116247 (48 strip pack-professional use), 3116239 (12 strip pack-patient self test), Non-U.S. Catalog Numbers 11937642190 [packaged 48 strips per box] and 11937634190 [packaged 12 strips per box].
- Status
- Terminated
- Initiated
- 2006-10-18
- Posted
- 2006-11-30
- Terminated
- 2007-06-20
- Reason
- Erroneous Test Results: Monitor may display 'error' message or report falsely elevated patient results caused by insufficient amounts of active ingredient (thromboplastin) in the test strips.
- Root cause
- Other
- Action
- A press release was issued by the firm on 10/19/06. The recall letter dated 10/19/06 sent to distributors who sell to home users instructs them to notify their home users and to discard any of product 3116239 in stock. The recall letter dated 10/19/06 to to home users, instructs them to discontinue use of, and to discard, their test strips and to contact their health care professional. The device correction letter dated 10/19/06 to distributors to professional users instructs them to notify customers and provides instructions for duplicate testing in laboratories. The device correction letter dated 10/19/06 to health care professionals asks that they follow-up with their home users to ensure they follow the instructions to cease home use and provides instructions for dupicate testing in laboratories.
- Quantity
- Approx 26,000,000 test strips to U.S. and 31,000,000 test strips Ex-U.S.
- Distribution
- Worldwide
- Lot, serial or model codes
- All lots.
- 510(k) numbers
- ["K930454","K962571","K974569","K994349"]
- PMA numbers
- not on file
- Product code
- GJS
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06