Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific myLUX Patient Application for an Apple Mobile device, Model Number 2939, Version 2.0.101, 2.0.110, cardiac monitor application.

Recall number
Z-0226-2025
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific myLUX Patient Application for an Apple Mobile device, Model Number 2939, Version 2.0.101, 2.0.110, cardiac monitor application.
Status
Open, Classified
Initiated
2024-09-06
Posted
2024-10-25
Terminated
not on file
Reason
Patients using the Model 2939 myLUX Patient Application on their Apple mobile device (e.g., iPhone) may not be transmitting data from their Insertable Cardiac Monitor to the LATITUDE Clarity website.
Root cause
Under Investigation by firm
Action
A field action/communication with Health Care Providers (HCPs) for affected patients was initiated on September 6, 2024, to inform them of the issue and provide the necessary patient data and connection support. Clinicians and patients were contacted via telephone calls. The BSC RhythmCARE team first called clinicians to describe the monitoring problem. With permission of the clinician, RhythmCARE Support contacted the patient to request a patient-initiated interrogation of the insertable cardiac monitor. Next, they offered options to troubleshoot and resolve the monitoring problem. In most cases, the troubleshooting included providing a BSC provided mobile monitor, in a singular case, the patients chose to buy a new mobile phone. The goal of these troubleshooting calls was to re-establish patient monitoring as scheduled. Recommendations were provided on an individual basis
Quantity
7,649 apps
Distribution
Nationwide distribution to AL, AZ, CA, CO, FL, GA, ID, KY, MA, MD, NC, NJ, NY, OH, Puerto Rico, SC, UT, VA, WA.
Lot, serial or model codes
GTIN 00802526618215, Versions 2.0.101, 2.0.110
510(k) numbers
["K231328"]
PMA numbers
not on file
Product code
MXD
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register