FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).
- Recall number
- Z-0226-2015
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).
- Status
- Terminated
- Initiated
- 2014-10-23
- Posted
- 2014-11-26
- Terminated
- 2015-08-10
- Reason
- Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.
- Root cause
- Process control
- Action
- DePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated October 23, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check inventory and remove affected stock, call Synthes to obtain a Return Authorization for the device, and to return the completed verification section of the response form by fax to (610) 430-7083 or Scan/email: Fieldaction@synthes.com. Customers with questions were instructed to call 610-719-5450.
- Quantity
- 24
- Distribution
- Distributed to MI, MT, and CO.
- Lot, serial or model codes
- Part Number 04.634.002 Lot Number 6816781
- 510(k) numbers
- ["K100952"]
- PMA numbers
- not on file
- Product code
- NKB
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as