Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical PM, Inc.: Advisor Vital Signs Monitor (model 9200), catalog number 925454220, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main boards (P/N: 80002B1) CE/MDD marking, 3-lead 50 Hz ECG and standard SPO2 option.

Recall number
Z-0226-06
Recalling firm
Smiths Medical PM, Inc.
Product
Advisor Vital Signs Monitor (model 9200), catalog number 925454220, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main boards (P/N: 80002B1) CE/MDD marking, 3-lead 50 Hz ECG and standard SPO2 option.
Status
Terminated
Initiated
2005-09-01
Posted
2005-11-30
Terminated
2007-06-27
Reason
Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
Root cause
Other
Action
Distirbutors and consignees were notified beginning 09/01/05 via email or phone about Safety Action Bulletin 05-004. Customers were asked to respond to Smiths Medical with a written acknowledgement that they received this Safety Action Bulletin (SAB). The SAB described the issue, actions they should take and instruction on how to have the monitor repaired.
Quantity
1
Distribution
AZ, CA, KS, KY, NY, TN, TX OUS to countries such as: Egypt, France, Iran, Mexico, Peru, Romania, Saudi Arabia and Spain
Lot, serial or model codes
serial number: AM05070061
510(k) numbers
["K010770","K011177","K030407","K982279"]
PMA numbers
not on file
Product code
DPS
Firm FEI
2183502
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register