Skip to the content

FDA Medical Device Recalls (openFDA)

UIH Technologies LLC: Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only

Recall number
Z-0223-2026
Recalling firm
UIH Technologies LLC
Product
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only
Status
Open, Classified
Initiated
2025-09-12
Posted
2025-10-21
Terminated
not on file
Reason
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Root cause
Other
Action
On September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions: 1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts. 2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers. 3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799 4.Report adverse events: if any adverse events or quality issue occur, please report them to the FDA via the MedWatch program. 5. Please share this information with all relevant personnel at your facility, including technologist, service staff, and clinical teams.
Quantity
95 units
Distribution
US: OUS:
Lot, serial or model codes
UDI: (01)06971576832026/ Serial Numbers: 200017 200023 200024 200036 200045 200047 230002 230003 230004 230005 230006 230007 230008 230009 230011 230010 230012 230014 230013 200059 230015 230016 230017 230019 230020 230021 230022 230023 230024 230025 232001 200071 230026 230027 200072 230028 200074 230030 230029 200075 200076 230031 230032 230033 230034 230035 200079 200078 230037 230036 230038 230039 230040 230041 230042 230043 230044 230045 230047 230048 232002 202002 232003 232004 232005 230057 232006 232007 202004 232008 232009 202005 232010 202006 232011 202007 232012 232013 232014 230067 230068 230069 232015 232016 230070 230071 230073 202008 230072 230074 232017 202009 232018 232019 202010
510(k) numbers
["K182237","K193241"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3014809646
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register