Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Medical Corp: Glidesheath Slender Hydrophilic Coated Introducer Sheath, 5FR - 10CM (0.021), REF/Product Code RM*RS5F10PA, STERILE, Rx ONLY

Recall number
Z-0223-2018
Recalling firm
Terumo Medical Corp
Product
Glidesheath Slender Hydrophilic Coated Introducer Sheath, 5FR - 10CM (0.021), REF/Product Code RM*RS5F10PA, STERILE, Rx ONLY
Status
Terminated
Initiated
2017-06-07
Posted
not on file
Terminated
2019-04-25
Reason
An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath.
Root cause
Nonconforming Material/Component
Action
URGENT VOLUNTARY MEDICAL DEVICE RECALL Letters (dated June 7, 2017) and Response Forms were sent to the consignees via UPS on June 7, 2017.
Quantity
not on file
Distribution
Nationwide and Canada
Lot, serial or model codes
Lot s 161019, 161115, 161212, 170112
510(k) numbers
["K142183"]
PMA numbers
not on file
Product code
DYB
Firm FEI
2243441
Firm city
Somerset
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register