Skip to the content

FDA Medical Device Recalls (openFDA)

Dako North America Inc.: EnVision FLEX/HRP visualization reagent found in the following kits: EnVision FLEX, High pH (Link) EnVision FLEX+, Mouse, High pH (Link) EnVision FLEX Mini Kit, High pH (Link) EnVision FLEX, High pH (Dako Omnis) Multi-component, sealed package. For in vitro diagnostic use.

Recall number
Z-0220-2016
Recalling firm
Dako North America Inc.
Product
EnVision FLEX/HRP visualization reagent found in the following kits: EnVision FLEX, High pH (Link) EnVision FLEX+, Mouse, High pH (Link) EnVision FLEX Mini Kit, High pH (Link) EnVision FLEX, High pH (Dako Omnis) Multi-component, sealed package. For in vitro diagnostic use.
Status
Terminated
Initiated
2015-09-04
Posted
2015-11-05
Terminated
2016-04-25
Reason
There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.
Root cause
Other
Action
Dako sent an customer notification letter on September 4, 2015, to all affected customers via mail. The letter identified the product the problem and the action needed to be taken by the customer. The customers/distributors are instructed in the Recall Notification Letters to 1) remove any unused kits bearing the affected lot numbers listed in their letter from diagnostic use 2) destroy any unused kits bearing the affect lot numbers listed in their letter, and 3) determine any possible impact of this change in performance on patient results from the affected lots. The letter states actions should be taken within 10 days and that customers/distributors should confirm they have received the information by returning the attached, fully completed Device Recall Form and return it to Dako QA Vigilance by email at dako.dkvigilance@dako.com. The firm lists on their letter for customers to contact their Dako representative in there are any questions. For further questions, call (805) 566-6655.
Quantity
300 units
Distribution
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and to the states of : TX, SD, CA, FL, MA, MD, WV, NC, OH, DE, MI, NJ, CO, HI, AR, NY, LA, GA, NV, RI, IL, OK, CT, MO, VA, and DC., and to the countries of : Canada, Brazil and Costa Rica.
Lot, serial or model codes
Item No.: K800021-2, K800221-2, K802321-2 (Link) K801021-2, K801221-2, K802421-2 (AS/AS+) GV80011-2, GV82311-2 (Dako Omnis)  Catalog/Model No.: K8000 K8002 K8023 GV800  SM802 EnVision FLEX / HRP Lot 20019374 used in: Envision FLEX, High pH (link) Lot No.: 20019097, 20020777, 20021381 Envision FLEX+, Mouse, High pH (link) Lot No.: 20019103, 20019113, 20020772 EnVision FLEX Mini Kit, High pH (Link) Lot No.: 20021362  DM 842 EnVision FLEX / HRP (Dako Omnis) Lot 20019013 used in: EnVision FLEX, High pH (Dako Omnis) Lot No.: 20019102
510(k) numbers
not on file
PMA numbers
not on file
Product code
LDT
Firm FEI
2022180
Firm city
Carpinteria
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register