Skip to the content

FDA Medical Device Recalls (openFDA)

General Electric Med Systems LLC: Precision 500D Radiographic/Fluoroscopic X-ray System, model numbers 2225849, 005853. The Precision 500D system consists of an X-ray generator; angulating table with X-ray Tube, collimator and image intensifier; wall stand; overhead tube suspension; operator console and digital archive system.

Recall number
Z-0220-05
Recalling firm
General Electric Med Systems LLC
Product
Precision 500D Radiographic/Fluoroscopic X-ray System, model numbers 2225849, 005853. The Precision 500D system consists of an X-ray generator; angulating table with X-ray Tube, collimator and image intensifier; wall stand; overhead tube suspension; operator console and digital archive system.
Status
Terminated
Initiated
2004-12-30
Posted
2006-02-14
Terminated
2006-02-21
Reason
Units have a defect in accordance with 21 CFR 1003.2 because an incorrect exposure can occur.
Root cause
Other
Action
An 'Urgent: Radiation Safety Notice' letter was issued to hospitals with affected systems describing potential hazards, the planned modification and recommended action to limit the hazard until the planned modification is installed. A GE field Service Representative will install the modification.
Quantity
921
Distribution
globally
Lot, serial or model codes
all serial numbers
510(k) numbers
["K011624"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register