FDA Medical Device Recalls (openFDA)
General Electric Med Systems LLC: Precision 500D Radiographic/Fluoroscopic X-ray System, model numbers 2225849, 005853. The Precision 500D system consists of an X-ray generator; angulating table with X-ray Tube, collimator and image intensifier; wall stand; overhead tube suspension; operator console and digital archive system.
- Recall number
- Z-0220-05
- Recalling firm
- General Electric Med Systems LLC
- Product
- Precision 500D Radiographic/Fluoroscopic X-ray System, model numbers 2225849, 005853. The Precision 500D system consists of an X-ray generator; angulating table with X-ray Tube, collimator and image intensifier; wall stand; overhead tube suspension; operator console and digital archive system.
- Status
- Terminated
- Initiated
- 2004-12-30
- Posted
- 2006-02-14
- Terminated
- 2006-02-21
- Reason
- Units have a defect in accordance with 21 CFR 1003.2 because an incorrect exposure can occur.
- Root cause
- Other
- Action
- An 'Urgent: Radiation Safety Notice' letter was issued to hospitals with affected systems describing potential hazards, the planned modification and recommended action to limit the hazard until the planned modification is installed. A GE field Service Representative will install the modification.
- Quantity
- 921
- Distribution
- globally
- Lot, serial or model codes
- all serial numbers
- 510(k) numbers
- ["K011624"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28