FDA Medical Device Recalls (openFDA)
AGFA Corp.: CR NX Modality Workstation, Computed Radiography System (Software versions NX 1.0.2402 and NX 1.0.2405).
- Recall number
- Z-0219-2007
- Recalling firm
- AGFA Corp.
- Product
- CR NX Modality Workstation, Computed Radiography System (Software versions NX 1.0.2402 and NX 1.0.2405).
- Status
- Terminated
- Initiated
- 2006-09-01
- Posted
- 2006-11-28
- Terminated
- 2008-04-24
- Reason
- Users may experience one or more of four (4) problems; 1. Image quality problem with full leg full spine exams. 2. Unable to print 2 or more images on 1 film. 3. Possible image loss when printing. 4. Study date and time not showing up on Imexius Web.
- Root cause
- Other
- Action
- A Service Bulletin to address the issues was created on 09/01/2006. Customers were notified of the plans to upgrade via telephone and email on 10/13/2006.
- Quantity
- 178 units
- Distribution
- Nationwide
- Lot, serial or model codes
- NX Workstation Software is sold in the following packages: 60+0026406, 60+00026407, 60+00026408, 60+00026857, 60+00026858, 60+00026410, ES885. Software versions: NX 1.0.2402 and NX 1.0.2405
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MBQ
- Firm FEI
- 3001236302
- Firm city
- Greenville
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28