FDA Medical Device Recalls (openFDA)
Gebauer Company: Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The product consists of a can, a valve and an actuator. Gebauer's Fluro-Ethyl is a vapocoolant (skin refrigerant), topical anesthetic intended to control the pain associated with minor surgical procedures, dermabrasion and injection. It is also effective in providing temporary relief from the pain associated with minor sports injuries.
- Recall number
- Z-0218-2009
- Recalling firm
- Gebauer Company
- Product
- Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The product consists of a can, a valve and an actuator. Gebauer's Fluro-Ethyl is a vapocoolant (skin refrigerant), topical anesthetic intended to control the pain associated with minor surgical procedures, dermabrasion and injection. It is also effective in providing temporary relief from the pain associated with minor sports injuries.
- Status
- Terminated
- Initiated
- 2008-09-03
- Posted
- 2008-11-12
- Terminated
- 2010-10-13
- Reason
- Some units of Fluro-Ethyl containing the defective valve which could malfunction and spray refrigerant out from the side of the valve in addition to or instead of spraying in the normal inverted position from the product's actuator. The product can also leak coolant from the side of the valve onto the fingers of the user.
- Root cause
- Component design/selection
- Action
- A recall notification letter (Recall Notice) (dated 9/3/2008) was issued on 9/16/2008 to customers. The notice instructs customers to check their inventory for any remaining recall product and if found, return the product to the Gebauer Company. The letter also advises distributors to notify their customers and follow instructions as noted above, i.e. customers inventory their stock for recalled product and if found, return to the Gauber Company. For questions or additional information, contact Gebauer Company at 216-581-3030 (ext. 120) or 800-321-9348.
- Quantity
- 12,717 Units
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Model Number: P/N 0386-0002-20. Lot Numbers: 7010, 7011, and 7012.
- 510(k) numbers
- ["K002596"]
- PMA numbers
- not on file
- Product code
- MLY
- Firm FEI
- 1519179
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28