Skip to the content

FDA Medical Device Recalls (openFDA)

Gebauer Company: Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The product consists of a can, a valve and an actuator. Gebauer's Fluro-Ethyl is a vapocoolant (skin refrigerant), topical anesthetic intended to control the pain associated with minor surgical procedures, dermabrasion and injection. It is also effective in providing temporary relief from the pain associated with minor sports injuries.

Recall number
Z-0218-2009
Recalling firm
Gebauer Company
Product
Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The product consists of a can, a valve and an actuator. Gebauer's Fluro-Ethyl is a vapocoolant (skin refrigerant), topical anesthetic intended to control the pain associated with minor surgical procedures, dermabrasion and injection. It is also effective in providing temporary relief from the pain associated with minor sports injuries.
Status
Terminated
Initiated
2008-09-03
Posted
2008-11-12
Terminated
2010-10-13
Reason
Some units of Fluro-Ethyl containing the defective valve which could malfunction and spray refrigerant out from the side of the valve in addition to or instead of spraying in the normal inverted position from the product's actuator. The product can also leak coolant from the side of the valve onto the fingers of the user.
Root cause
Component design/selection
Action
A recall notification letter (Recall Notice) (dated 9/3/2008) was issued on 9/16/2008 to customers. The notice instructs customers to check their inventory for any remaining recall product and if found, return the product to the Gebauer Company. The letter also advises distributors to notify their customers and follow instructions as noted above, i.e. customers inventory their stock for recalled product and if found, return to the Gauber Company. For questions or additional information, contact Gebauer Company at 216-581-3030 (ext. 120) or 800-321-9348.
Quantity
12,717 Units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Model Number: P/N 0386-0002-20. Lot Numbers: 7010, 7011, and 7012.
510(k) numbers
["K002596"]
PMA numbers
not on file
Product code
MLY
Firm FEI
1519179
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register