Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Health Products, Inc.: IMx HAVAB Controls are for In Vitro Diagnostic use. The kit is composed of 2 bottles of Negative and Positive controls of 9 mL each. These are prepared with recalcified human plasma and the preservative is Sodium Azide. The Negative Control is non-reactive for HbsAG, HIV-1, Anti-HCV, anti-HiV-1/HIV-2 and IgG antibody against HAV. The Positive Control is recalcified human plasma for anti-HAV diluted with Negative Control.

Recall number
Z-0217-05
Recalling firm
Abbott Health Products, Inc.
Product
IMx HAVAB Controls are for In Vitro Diagnostic use. The kit is composed of 2 bottles of Negative and Positive controls of 9 mL each. These are prepared with recalcified human plasma and the preservative is Sodium Azide. The Negative Control is non-reactive for HbsAG, HIV-1, Anti-HCV, anti-HiV-1/HIV-2 and IgG antibody against HAV. The Positive Control is recalcified human plasma for anti-HAV diluted with Negative Control.
Status
Terminated
Initiated
2004-10-27
Posted
2004-11-20
Terminated
2005-03-16
Reason
Abbott identified through investigational studies that IMx HAVAB Controls lot 18220Q100 is generating Negative Control values outside the upper range specified in the IMx HAVAB Package Insert.
Root cause
Other
Action
Device recall letters were dated 10/27/2004. Letter was mail to customers asking to destroy and discontinue use of device. On hand inventory information was requested to be forward to Abbott at 1-800-777-0051 (U.S. only).
Quantity
397 kits
Distribution
The product has been distributed to hospitals, laboratories and other institution including government accounts. Geographic areas of distribution include Puerto Rico, Canada, Germany, Thailand and U.S.A. Government accounts includes: VA Medical Center, Saint Louise, MO; Brecksville VA Hosp. Supp, Brecksville, OH; VA MED CENTER, Phoenix, AZ; Department Veterinary Affair Med., Houston, TX; VA Med., Cincinnati, OH; VA Med Center, Seattle, WA; Truman Mem. VA, Columbia, MO; Dept. Vet. Aff., Shreveport, LA; VA Hosp. Whse., Fresno, CA ; VA Med Center, Omaha, NE ; VA Med. Center, Lexington, KY; VA Med Center, Detroit, MI ; VA Hosp., San Diego, CA ; Naval Hosp., Great Lakes, IL; VA Med. Center, San Francisco, CA.
Lot, serial or model codes
Lot # 18220Q100 Exp. March 02, 2005
510(k) numbers
not on file
PMA numbers
["P780012S007"]
Product code
LOL
Firm FEI
2623532
Firm city
Barceloneta
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register