Skip to the content

FDA Medical Device Recalls (openFDA)

ELITech Clinical Systems SAS: ELITech Clinical Systems reagents used on the Selectra ProS and ProM systems that utilize the Trinder Reaction method. The following product reference numbers and names: GPSL-025O: Glucose PAP SL AUML-0250: URIC ACID MONO SL CHSL-0250: CHOLESTEROL SL, TGML-0250: TRIGLYCERIDES SL, CRSL-0250: CREATININE PAP SL HDLL-0230: CHOLESTEROL HDL SL 2G LDLL-0230: CHOLESTEROL LDL SL 2G

Recall number
Z-0216-2016
Recalling firm
ELITech Clinical Systems SAS
Product
ELITech Clinical Systems reagents used on the Selectra ProS and ProM systems that utilize the Trinder Reaction method. The following product reference numbers and names: GPSL-025O: Glucose PAP SL AUML-0250: URIC ACID MONO SL CHSL-0250: CHOLESTEROL SL, TGML-0250: TRIGLYCERIDES SL, CRSL-0250: CREATININE PAP SL HDLL-0230: CHOLESTEROL HDL SL 2G LDLL-0230: CHOLESTEROL LDL SL 2G
Status
Terminated
Initiated
2015-09-22
Posted
2015-11-05
Terminated
2016-03-04
Reason
ELITech Clinical Systems reagents based on Trinder reaction used on the Selectra ProS and ProM systems are recalled due to potential drug interference. No complaints were received.
Root cause
Under Investigation by firm
Action
The firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to customercare@elitechgroup.com or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
Quantity
969 systems
Distribution
US Nationwide Distribution
Lot, serial or model codes
all lots
510(k) numbers
["K100525"]
PMA numbers
not on file
Product code
CGA
Firm FEI
3007662974
Firm city
SEES
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register