Greiner Bio-One North America, Inc.: VACUETTE, 2 ml Lithium Heparin Venous Blood Collection Tubes. Reference #454237, 13x75 green cap-white ring, 24 racks of 50 pcs (1200 pcs in total, non ridged, Sterile). Manufactured by: Greiner bio-one NA, Inc., 4238 Capital Drive, Monroe, NC 28110. The Greiner Vacuette blood collection tube with lithium heparin and gel separator is an evacuated blood collection device containing lithium heparin, an anticoagulant additive, and an inert polymeric barrier material. The product is intended for use in holding and separating blood plasma from the cellular components of blood.

Recall number
Z-0215-2009
Recalling firm
Greiner Bio-One North America, Inc.
Product
VACUETTE, 2 ml Lithium Heparin Venous Blood Collection Tubes. Reference #454237, 13x75 green cap-white ring, 24 racks of 50 pcs (1200 pcs in total, non ridged, Sterile). Manufactured by: Greiner bio-one NA, Inc., 4238 Capital Drive, Monroe, NC 28110. The Greiner Vacuette blood collection tube with lithium heparin and gel separator is an evacuated blood collection device containing lithium heparin, an anticoagulant additive, and an inert polymeric barrier material. The product is intended for use in holding and separating blood plasma from the cellular components of blood.
Status
Terminated
Initiated
2008-07-29
Posted
2008-11-07
Terminated
2009-08-26
Reason
Labeling Error: Incorrect tube label reads "Z Serum Clot Activator" instead of "LH Lithium Heparin" tube. All other information on the tube label including color coding of tube was correct. Case and rack label information showed the correct information with 2 ml Lithium Heparin.
Root cause
Labeling mix-ups
Action
Consignees were notified by letter on 07/29/2008. Notification letters to distributors instructed them to stop distribution of the product and destroy all affected lots of the product. In addition, notification letters addressed to users instructed them to destroy and discard all products from the item and affected lots. The "Product Disposition Form" must be completed by the consignee and returned to Greiner Bio-One North America, Inc. For questions or additional information, contact Greiner Bio-One North America, Inc. at 704-261-7800.
Quantity
33 cases (39,600 pieces)
Distribution
Worldwide including USA (states of AK, AL, AR, CA, CO, DE, GA, ID, IL, IN, KS, LA, MA, MI, MO, MS, NC, ND, NE, NH, NY, OH, OK, OR, PA, RI, SC, SD, TX, VA, VT, WA, and WI) and country of Canada.
Lot, serial or model codes
Lot # B120703.
510(k) numbers
["K960857"]
PMA numbers
not on file
Product code
JKA
Firm FEI
3001451379
Firm city
Monroe
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register