Stryker Howmedica Osteonics Corp.: Stryker T2 Locking Screw, Fully threaded; 05 x 70mm, Catalog Number: 1896-5070S, Sterile, Qty: 1,Stryker Trauma GMbH, 24232 Schonkirchen, Germany Distributed in the United States by Howmedica Osteonics Corp.; 325 Corporate Drive, Mahwah, NJ 07430

Recall number
Z-0214-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker T2 Locking Screw, Fully threaded; 05 x 70mm, Catalog Number: 1896-5070S, Sterile, Qty: 1,Stryker Trauma GMbH, 24232 Schonkirchen, Germany Distributed in the United States by Howmedica Osteonics Corp.; 325 Corporate Drive, Mahwah, NJ 07430
Status
Terminated
Initiated
2007-09-27
Posted
2007-11-16
Terminated
2008-05-08
Reason
Mislabeled: One lot of screws (05x70mm) is labeled as containing 70mm, but may actually contain 75mm length screws.
Root cause
Packaging process control
Action
Urgent Product Recall letters were sent on September 27, 2007, to all Stryker Branches/Agencies and hospitals who received the product requesting them to remove and quarantine the product for Stryker pick-up.
Quantity
50
Distribution
Nationwide
Lot, serial or model codes
Catalog number: 189605970S: Lot Code K871252
510(k) numbers
["K003018"]
PMA numbers
not on file
Product code
not on file
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register