Skip to the content

FDA Medical Device Recalls (openFDA)

Sterilmed Inc: St. Jude Medical. Agilis NxT Steerable Introducer: B1-Directional, Medium Curl 22.4 mm, 71 cm, 6.5 F ID; Dilator: 94 cm, 8.5 F; Super Stiff, Guidewire: 180 cm, 0.032". Reprocessed device for single use only, sterile. Product Usage: The reprocessed steerable introducer sheaths are indicated when introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

Recall number
Z-0213-2018
Recalling firm
Sterilmed Inc
Product
St. Jude Medical. Agilis NxT Steerable Introducer: B1-Directional, Medium Curl 22.4 mm, 71 cm, 6.5 F ID; Dilator: 94 cm, 8.5 F; Super Stiff, Guidewire: 180 cm, 0.032". Reprocessed device for single use only, sterile. Product Usage: The reprocessed steerable introducer sheaths are indicated when introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Status
Terminated
Initiated
2017-06-02
Posted
2017-12-29
Terminated
2019-07-29
Reason
Inadequate seal due to lack of adhesive glue
Root cause
Process control
Action
Sterilmed sent an Urgent Medical Device Recall letter dated June 12, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory to determine if affected products are on hand, complete and return the Business Reply Form, share information with appropriate staff in facility, attach copy of this notification to quarantined product, and return affected product to Sterilmed. For questions contact Sterilmed's Customer Care Team at 888-541-0078.
Quantity
112 devices
Distribution
US Nationwide distribution in the states of AR, CO, FL, MI, OH, TX, and WA.
Lot, serial or model codes
Model No. STJ408309, STJ408310, STJ408324. Exp. Date of 05/31/2018 or PRIOR. All Lots, known lots: 7344024, 7343120, 7343643, 7343797, 7348305, 7355750, 7348712, 7290257, 7354522, 7354523, 7350001, 7324499, 7343501, 7351203, 7255673, 7273483, 7343603, 7355439, 7355745, 7355754, 7355752, 7282741, 7315539, 7320863, 7343766, 7351038, 7330645, 7345143, 7323722, 7330603, 7342081, 7351724, 7352687, 7355130, 7355131, 7355137, 7377760, 7355135, 7375662, 7404300, 7404301, 7404316, 7404306, 7343395, 7355463, 7355714, 7355622, 7330666, 7355572, 7327999, 7355677, 7308337, 7343686, 7355680, 7251903, 7259400, 7322815, 7342069, 7355574, 7355676, 7355712, 7346336, 7355666, 7355661, 7355618, 7355571, 7354437, 7350573, 7331054, 7300290, 7350792, 7355679, 7355709, 7355713, 7355681, 7355675, 7355573, 7345129, 7343838, 7355678, 7355575, 7355620, 7355711, 7346820, 7355662, 7404299, 7404282, 7404294, 7404290, 7404310, 7404311, 7404318, 7404297, 7404287, 7342672, 7357129, 7241712, 7350942, 7272027, 7316930, 7345592, 7348169, 7388731, 7372138, 7386929, 7387147
510(k) numbers
["K152090"]
PMA numbers
not on file
Product code
DYB
Firm FEI
3011276773
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register