FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corporation: HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974
- Recall number
- Z-0212-2024
- Recalling firm
- Baxter Healthcare Corporation
- Product
- HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974
- Status
- Open, Classified
- Initiated
- 2023-10-04
- Posted
- 2023-11-02
- Terminated
- not on file
- Reason
- Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.
- Root cause
- Under Investigation by firm
- Action
- Baxter issued an Urgent Medical Device Correction notice to its consignees on 10/04/2023 by USPS first class mail. The notice explained the potential problem and hazard. Baxter will be correcting this issue by installing a conductive drag chain on the Progressa+ bed frame to prevent the electrostatic charge buildup. A Baxter representative will contact your facility to schedule a device correction to add the conductive drag chain. Please note you will be receiving this correction from Baxter at no charge. All units purchased in the future will have the drag chain.
- Quantity
- 499 beds
- Distribution
- US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA
- Lot, serial or model codes
- UDI/DI 00887761966925, all serial numbers
- 510(k) numbers
- ["K122473"]
- PMA numbers
- not on file
- Product code
- IOQ
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28