AGFA Corp.: IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.

Recall number
Z-0211-2012
Recalling firm
AGFA Corp.
Product
IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.
Status
Terminated
Initiated
2011-06-06
Posted
2011-11-18
Terminated
2013-07-02
Reason
Improper merging of IMPAX Cardiovascular Results Management (RM) reports causes the signed report to become unavailable.
Root cause
Software design
Action
AGFA Healthcare sent an "URGENT FIELD SAFETY NOTICE" letter dated May 27, 2011, via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to carry out a merge post finalization of an RM reports. If it is necessary to modify a finalized report, AGFA recommends that they make the modification through one of the following actions: Amend the report through the RM interface, or Call AGFA Service to change the status of the report to "preliminary" and then proceed with the finalization of the report. AGFA issued an Acknowledgment, via FAX- Back that the information was received and understood to affected customers. For any questions customers were instructed to contact their AGFA Healthcare Service at 877-777-2432.
Quantity
323
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Lot, serial or model codes
Software Versions: RM2.04.37.04 to RM 7.8 SU2 (RM 2.04.37.04 is a version prior to CV 7.2 release).
510(k) numbers
["K050228"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register