Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: AMSCO 400 and AMSCO C Small Steam Sterilizers

Recall number
Z-0210-2016
Recalling firm
Steris Corporation
Product
AMSCO 400 and AMSCO C Small Steam Sterilizers
Status
Terminated
Initiated
2015-09-03
Posted
2015-11-04
Terminated
2016-01-22
Reason
In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close the sterilizer door.
Root cause
Software design
Action
On 9/21/2015 the firm sent Recall Correction Notices to their customers.
Quantity
32 units
Distribution
US nationwide distribution.
Lot, serial or model codes
Serial #'s: 033331205, 033341248, 033351210, 033401221, 033411215, 033451217, 030741306, 031491314, 032381302, 032341304, 033281222, 030621412, 030871412, 030871410, 031751401, 031771425, 031781412, 031781414, 031911436, R020141502, R020151501, R020191502, R020201501, R020261502, R020261504, R020271501, 030341514, R020471501, 030501540, 030651519, R021001503, R021051501
510(k) numbers
["K111223"]
PMA numbers
not on file
Product code
FLE
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register