Skip to the content

FDA Medical Device Recalls (openFDA)

Cyberonics, Inc: VNS Therapy Aspire HC Generator and VNS, Model 105 Cyberonics, Inc. 100 Cyberonics Blvd., Houston, TX 77058 For implantation in humans to treat Epilepsy and Depression.

Recall number
Z-0210-2012
Recalling firm
Cyberonics, Inc
Product
VNS Therapy Aspire HC Generator and VNS, Model 105 Cyberonics, Inc. 100 Cyberonics Blvd., Houston, TX 77058 For implantation in humans to treat Epilepsy and Depression.
Status
Terminated
Initiated
2011-08-12
Posted
2011-11-18
Terminated
2012-08-20
Reason
The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.
Root cause
Nonconforming Material/Component
Action
Cyberonics sent a Field Safety Alert on August 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The firm decided to recall and initiated contact with implanting facilities in the United States on 08/15/2011. A Field Safety Alert was delivered on 09/19/2011 to implanting facilities. Customers were instructed not to prescribe AspireHC pulse generators for implant until Cyberonics resumes shipment of AspireHC pulse generators that have the issue remedied. Cyberonics recommends that physicians continue their current practice of titrating programming parameters for patients implanted with an AspireHC pulse generator to achieve appropriate clinical outcomes, paying particular attention to report of: Lack or loss of efficacy Pain, inflammation, swelling/edema in the neck area; and/or Patient perception of stimulation or muscle twitching in the area surrounding the implant. Cyberonics recommended routine systems diagnostic testing to aid in device and patient monitoring. Physicians may contact Clinical Technical Support at 866-882-8804 or via email at clinicaltechnicalservices@cyberonics.com. For questions regarding this recall call 281-228-7323.
Quantity
340 units
Distribution
Worldwide Distribution - USA including TN, TX, IL, OH, FL, CA, CO, OR, AZ, NY, PA, MO, NJ, WI, OK, MI, ID, KY, IA, MA, NE, GA, LA, NY, and Norway.
Lot, serial or model codes
Model 105, VNS Therapy AspireHC Generator
510(k) numbers
not on file
PMA numbers
["P970003"]
Product code
LYJ
Firm FEI
1000117004
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register