Skip to the content

FDA Medical Device Recalls (openFDA)

Sechrist Industries Inc: Sechrist Monoplace Hyperbaric Chamber with Electronic Control Module Model 3600E/ER The Hyperbaric Chamber is intended to provide 100% oxygen to the patient at greater than ambient pressures

Recall number
Z-0209-2013
Recalling firm
Sechrist Industries Inc
Product
Sechrist Monoplace Hyperbaric Chamber with Electronic Control Module Model 3600E/ER The Hyperbaric Chamber is intended to provide 100% oxygen to the patient at greater than ambient pressures
Status
Terminated
Initiated
2008-08-28
Posted
2012-11-06
Terminated
2012-11-06
Reason
Sechrist conducted a field upgrade on the hyperbaulic chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of the chamber.
Root cause
Software design
Action
Sechrist Industries contacted consignees by phone to notify them of a software upgarde and to schedule appointments to perform the required field upgrades. An approximate summary of the phone notification is provided below. "I am contacting you to notify you of a field upgrade we are performing on our hyperbaric chambers equipped with Electronic Control Modules in order to correct a software bug that could potentially affect the performance of your chamber. The upgrade includes the installation of a new motherboard with revised software, a reset switch, and the exchange of other control panel components. The process will take approximately 2.5 hours and will be performed at no charge to you." For questions regarding this recall call 714-579-8400.
Quantity
65 units total (61 units in the US)
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Bulgaria, and Canada
Lot, serial or model codes
Serial Numbers: 360001-360049, 360051-360064, 360068.
510(k) numbers
["K052713"]
PMA numbers
not on file
Product code
CBF
Firm FEI
2020676
Firm city
Anaheim
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register