Skip to the content

FDA Medical Device Recalls (openFDA)

Intersurgical Inc: Solus Flexible wire-reinforced laryngeal mask airways under the following sizes: 1) small pediatric size 2 catalog #8002001, 2) large pediatric size 2.5 catalog #8025001, 3) small adult size 3 catalog #8003001, 4) medium adult size 4 catalog #8004001 and 5) large adult size 5 catalog #8005001 Patent airway management.

Recall number
Z-0208-2018
Recalling firm
Intersurgical Inc
Product
Solus Flexible wire-reinforced laryngeal mask airways under the following sizes: 1) small pediatric size 2 catalog #8002001, 2) large pediatric size 2.5 catalog #8025001, 3) small adult size 3 catalog #8003001, 4) medium adult size 4 catalog #8004001 and 5) large adult size 5 catalog #8005001 Patent airway management.
Status
Terminated
Initiated
2017-04-03
Posted
not on file
Terminated
2019-06-17
Reason
A defect in the wire-reinforced tubing may result in a partial of total occlusion when the cuff is over-inflated.
Root cause
Nonconforming Material/Component
Action
Intersurgical sent an Urgent Medical Device Recall Notice dated April 3rd, 2017, along with response forms to Customers. Customers were advised to immediately discontinue use and quarantine affected products. If affected products were further distributed, customers are advised to notify those individuals and advise them of the recall situation and have them return their outstanding stock to you. Customers with questions can call the IS Customer Service line at 800-828-9633 for a Return Authorization Number. Once customers receive the Return Authorization Number, please enter it in the space provided on the Recall Response Form. The completed Recall Response Form should be emailed to: Support@intersurgicalinc.com For further questions, please call (315) 451-2900)
Quantity
144 cases / 20 units
Distribution
Worldwide Distribution - US Distribution to the states of : IL, KY, OH, NY & VA and to the countries of : Canada & Mexico
Lot, serial or model codes
Product Lots: 31402822 to 31610791
510(k) numbers
not on file
PMA numbers
not on file
Product code
BYG
Firm FEI
1000120301
Firm city
E. Syracuse
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register