Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Model 801763.
- Recall number
- Z-0208-05
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Product
- Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Model 801763.
- Status
- Terminated
- Initiated
- 2003-08-12
- Posted
- 2004-11-18
- Terminated
- 2005-09-22
- Reason
- The central control monitor screen may go blank during use because of a bad wiring connection in the wiring harness or a low voltage setting on the potentiometer.
- Root cause
- Other
- Action
- Service representatives were sent a service bulletin dated 8/12/03 instructing them to correct units in the field bearing serial numbers between 20 and 109.
- Quantity
- not on file
- Distribution
- United States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Japan, Korea, Singapore and United Kingdom.
- Lot, serial or model codes
- All units bearing serial numbers 20 through 109.
- 510(k) numbers
- ["K022947"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28