Sunquest Information Systems, Inc.: Sunquest Laboratory versions 5.3, 5.4, 6.1, 6.2 and 6.3. Sunquest Laboratory Information System is comprised of several modules which are involved in aspects such as blood banking, collection management, specimen tracking, and use with microbiology and molecular diagnostics. Key capabilities include: 1) Autoverify results directly from analyzers and reference laboratories 2) Track tubes and containers from order to final disposition 3) Monitor key metrics such as turnaround time and utilization using management reports 4) Produce pathology results with Structured Reports 5) Suggest microbiology workups online according to defined SOPs and rules 6) All data entered into the system is accessible through audit trails and Ad Hoc Report tools 7) Any change to patient results (inquiry, reports) is logged and available for review

Recall number
Z-0207-2009
Recalling firm
Sunquest Information Systems, Inc.
Product
Sunquest Laboratory versions 5.3, 5.4, 6.1, 6.2 and 6.3. Sunquest Laboratory Information System is comprised of several modules which are involved in aspects such as blood banking, collection management, specimen tracking, and use with microbiology and molecular diagnostics. Key capabilities include: 1) Autoverify results directly from analyzers and reference laboratories 2) Track tubes and containers from order to final disposition 3) Monitor key metrics such as turnaround time and utilization using management reports 4) Produce pathology results with Structured Reports 5) Suggest microbiology workups online according to defined SOPs and rules 6) All data entered into the system is accessible through audit trails and Ad Hoc Report tools 7) Any change to patient results (inquiry, reports) is logged and available for review
Status
Terminated
Initiated
2008-04-17
Posted
2008-10-31
Terminated
2011-12-05
Reason
Sunquest has identified a problem with IGO interfaces when performing an Online File Cleanup (OFC) that can lead to the test results of one patient filing to another patient when the order codes are the same. Performing an OFC prior to all results being assigned an accession number may cause the patient results to remain in a temporary file where they can be incorrectly associated to the wrong pat
Root cause
Component design/selection
Action
Customers were sent a preliminary notice on 4/17/2008 followed by customer notice PSN-08-04 on 05/07/2008 via email. The notice advises that this notice is follow-up to the Preliminary notice distributed on 4/17/08 and applies to all clients who have an Instrument Generated Order (IGO) interface. Sunquest has identified a problem with IGO interfaces when performing an Online File Cleanup (OFC) that can lead to the test results of one patient filing to another patient when the order codes are the same. Performing an OFC prior to all results being assigned an accession number may cause the patient results to remain in a temporary file where they can be incorrectly associated to the wrong patient. When this occurs, Sunquest generated reports and inquiry contains the incorrect results. The letter provides the various steps that cause the issue to occur. Sites Potentially Affected by this Problem: Sunquest Laboratory using Instrument Generated Orders (IGO) Interface, Version 5.3 and later Action Sunquest Information Systems is taking to eliminate this Problem: Change Request 1-62VWLJ has been created to address this issue. This will be available as part of Sunquest Laboratory Instrument Build version 1.6.0023 and later. To request the software correction, customers can create a Service request via the Client support Webpage at www.sunquestinfo.com/supportweb. The notice also provides temporary procedural instructions: Effective immediately, do NOT perform Online File Cleanup (OFC) on any IGO instrument interface methods. -Manual OFC should not be performed using functions OFC, or Instrument Maintenance (IX or IXM), option 11 for IGO interfaces only. -If using function Instrument Maintenance (IX or IXM), option 1, 2, 2, Method, 2 (Automatic File Cleanup). This should be set to "N". Then stop and re-start the Instrument Results Processor. Note: Online file cleanup for non-IGO interfaces needs to be performed as usual. -Pending and Overdue Logs need to be monit
Quantity
419 clients had the affected code
Distribution
Worldwide Distribution --- including USA and countries of Canada, Bermuda, Ireland, United Kingdom, United Arab Emirates, and Saudi Arabia.
Lot, serial or model codes
Sunquest Laboratory versions 5.3, 5.4, 6.1, 6.2 and 6.3.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000306472
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register