FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.

Recall number
Z-0206-2025
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
Status
Open, Classified
Initiated
2024-09-24
Posted
2024-10-23
Terminated
not on file
Reason
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
Root cause
Under Investigation by firm
Action
Medtronic neuromodulation issued an URGENT MEDICAL DEVICE RECALL notice to its consignees of 09/24/2024 via letter. The notice explained the issue, potential risk, and requested the affected devices in stock be returned. The notice is to be shared with those to whom the device was distributed. If the device has been implanted, the consignee was directed to contact their Medtronic Representative to determine if there are additional steps that may establish communication with the INS. On 07-Oct-24 and 08-Oct-24, a Medtronic representative successfully performed an update using a manufacturing software tool to reestablish communication to the affected devices. Consignees were provided a consignee notification regarding this action.
Quantity
39 units
Distribution
International distribution to the countries of Germany, UK, and Israel.
Lot, serial or model codes
GTIN 00763000813468, Serial Numbers: NND504236H, NND504252H; GTIN 00763000813475, Serial Numbers: NND504165H, NND504169H, NND504174H, NND504181H, NND504208H, NND504212H, NND504215H, NND504216H, NND504217H, NND504218H, NND504221H, NND504222H, NND504223H, NND504224H, NND504229H, NND504230H, NND504232H, NND504234H, NND504235H, NND504243H, NND504255H, NND504156H, NND504157H, NND504159H, NND504160H; GTIN 00763000813529, Serial Numbers: NND504167H, NND504178H, NND504179H, NND504180H, NND504182H, NND504227H, NND504228H; GTIN A7630008135201, Serial Numbers: NND504167H, NND504178H, NND504180H; GTIN A7630008134701, Serial Numbers: NND504181H, NND504243H
510(k) numbers
not on file
PMA numbers
["P960009"]
Product code
MHY
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register