FDA Medical Device Recalls (openFDA)
Medtronic Neuromodulation: Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
- Recall number
- Z-0206-2025
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
- Status
- Open, Classified
- Initiated
- 2024-09-24
- Posted
- 2024-10-23
- Terminated
- not on file
- Reason
- A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
- Root cause
- Under Investigation by firm
- Action
- Medtronic neuromodulation issued an URGENT MEDICAL DEVICE RECALL notice to its consignees of 09/24/2024 via letter. The notice explained the issue, potential risk, and requested the affected devices in stock be returned. The notice is to be shared with those to whom the device was distributed. If the device has been implanted, the consignee was directed to contact their Medtronic Representative to determine if there are additional steps that may establish communication with the INS. On 07-Oct-24 and 08-Oct-24, a Medtronic representative successfully performed an update using a manufacturing software tool to reestablish communication to the affected devices. Consignees were provided a consignee notification regarding this action.
- Quantity
- 39 units
- Distribution
- International distribution to the countries of Germany, UK, and Israel.
- Lot, serial or model codes
- GTIN 00763000813468, Serial Numbers: NND504236H, NND504252H; GTIN 00763000813475, Serial Numbers: NND504165H, NND504169H, NND504174H, NND504181H, NND504208H, NND504212H, NND504215H, NND504216H, NND504217H, NND504218H, NND504221H, NND504222H, NND504223H, NND504224H, NND504229H, NND504230H, NND504232H, NND504234H, NND504235H, NND504243H, NND504255H, NND504156H, NND504157H, NND504159H, NND504160H; GTIN 00763000813529, Serial Numbers: NND504167H, NND504178H, NND504179H, NND504180H, NND504182H, NND504227H, NND504228H; GTIN A7630008135201, Serial Numbers: NND504167H, NND504178H, NND504180H; GTIN A7630008134701, Serial Numbers: NND504181H, NND504243H
- 510(k) numbers
- not on file
- PMA numbers
- ["P960009"]
- Product code
- MHY
- Firm FEI
- 2182207
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28