ConMed Corporation: Arthroscopic Energy 50 Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures.
- Recall number
- Z-0205-2016
- Recalling firm
- ConMed Corporation
- Product
- Arthroscopic Energy 50 Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures.
- Status
- Terminated
- Initiated
- 2015-06-17
- Posted
- 2015-11-03
- Terminated
- 2017-04-11
- Reason
- Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage.
- Root cause
- Software design (manufacturing process)
- Action
- On June 17, 2015 Con Med Corporation distributed Urgent Device Recall Notification Letters and Recall Reply forms to their customers via courier service. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory for any of the affected lot numbers referenced on the recall notice. If any of the recalled product are in stock, customers should discontinue the use or distribution of these items. If product was further distributed, customers should notify those individuals of this recall action. Customers with questions can call Gregory Connell at 727-399-5276 or email AES50S@conmed.com. Customers should also complete and return the Recall Reply Form by mail to the address listed on the recall notice.
- Quantity
- 276 devices
- Distribution
- Worldwide - US distribution in the states of FL, OH, and the country of Spain.
- Lot, serial or model codes
- 1412161, 1410201, 1410271, 1412121, and 1412081
- 510(k) numbers
- ["K140578"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 3001451571
- Firm city
- Utica
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28