Skip to the content

FDA Medical Device Recalls (openFDA)

Ormco/Sybronendo: Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.

Recall number
Z-0205-2013
Recalling firm
Ormco/Sybronendo
Product
Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.
Status
Terminated
Initiated
2011-12-12
Posted
2012-11-05
Terminated
2012-11-05
Reason
The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130C. The correct sterilization temperature should be listed as 132C and/or 270F.
Root cause
Error in labeling
Action
"Urgent - Medical Device Correction" letters dated 12/12/2011 were sent via USPS 1st class mail to all customers and again on 2/21/2012 for recall expansion. The device issue was described and recommended actions were provided. Customers were instructed to acknowledge and complete the Return Form. Questions were referred to (800) 854-1741.
Quantity
4,543
Distribution
Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique.
Lot, serial or model codes
Part Numbers 601-0007 and 601-0010, all lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
DZN
Firm FEI
2016150
Firm city
Glendora
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register